Instructions for use INTRA Motorhandstück PROPHYwiz 181 P - REF 1.003.2278
6 Reprocessing steps in accordance with ISO 17664 | 6.3 Automated reprocessing
14 / 22
NOTICE
Malfunctioning from using a disinfectant bath or chloride-containing
disinfectants.
Defects in the device.
▶
Disinfect manually only!
KaVo recommends the following products based on compatibility of the materi-
als. The microbiological efficacy must be ensured by the disinfectant manufac-
turer and proven by an expert opinion.
Approved disinfectants:
▪ Mikrozid AF made by Schülke & Mayr (liquid or cloths)
▪ FD 322 made by Dürr
▪ CaviCide made by Metrex
Consumables required:
▪ Cloths for wiping the medical device.
▶
Spray the disinfectant on a cloth, then wipe down the medical device and
allow the disinfectant to act according to the instructions of the
disinfectant manufacturer.
▶
Follow the instructions for use of the disinfectant.
6.2.4 Manual internal disinfection
Not applicable.
6.2.5 Manual drying
▶
If there is any residual moisture after cleaning and disinfecting, soak up the
residual moisture with a soft cotton cloth and then wipe dry.
▶
Immediately after drying, lubricate the KaVo medical device with care
agents from the KaVo care system.
See also:
2
6.4 Care products and systems - Servicing, Page 14
6.3 Automated reprocessing
Not applicable.
6.4 Care products and systems - Servicing
CAUTION
Improper service and care.
Risk of injury.
▶
Service regularly with suitable agents.