Instructions for use SONICflex 2000 N - 0.571.5020
6 Preparation methods according to ISO 17664 | 6.4 Drying
18 / 26
6.4 Drying
Manual Drying
▶ Blow off the outside and inside with compressed air until water drops are no lon‐
ger visible.
Automatic Drying
The drying procedure is normally part of the disinfection program of the thermodisin‐
fector.
Note
Please observe the instructions for use of the thermodisinfector.
6.5 Care products and systems - Servicing
CAUTION
Premature wear and malfunctions from improper servicing and care.
Reduced product life.
▶ Perform proper care regularly!
Note
KaVo only guarantees that its products will function properly when the care prod‐
ucts are used that are listed as accessories since they were tested for proper use
on our products.
Note
If you are bothered by some oil leaking during the treatment, the technical set-up
allows for switching from servicing with oil prior to sterilisation to once weekly serv‐
icing.
6.5.1 Torque wrench care
CAUTION
Malfunctions from cleaning in the ultrasonic unit.
Defects in the torque wrench.
▶ Do not place the torque wrench in ultrasonic cleaning devices.
If the torque wrench runs roughly, it should be lubricated with silicone grease (Mat.
no. 1.000.6403). The silicone grease is pressed into the torque wrench in the slots or
grease pockets of the locking springs. Then place a small amount of grease on your
fingertip, and press it into the torque wrench at the indicated location (see arrow).
Then rotate the torque wrench and regrease if necessary.
6.5.2 Care with KAVOspray
▶ Remove tip.
▶ Cover the product with the Cleanpac bag.