KaVo 0.571.5020 Instructions For Use Manual Download Page 1

Instructions for use

SONICflex 2000 N - 0.571.5020

Always be on the safe side.

Summary of Contents for 0.571.5020

Page 1: ...Instructions for use SONICflex 2000 N 0 571 5020 Always be on the safe side...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...Inserting the SONICflex tip 13 5 4 Removing the SONICflex tip 14 5 5 Regulating the spray 14 5 6 Check the amount of water 14 6 Preparation methods according to ISO 17664 15 6 1 Preparations at the s...

Page 4: ...6 8 Storage 20 7 Tools 21 Instructions for use SONICflex 2000 N 0 571 5020 Table of contents 4 26...

Page 5: ...d processing for a period of 12 months from the date of the invoice subject to the following conditions In case of justified complaints KaVo will honour its warranty with a free replacement or repair...

Page 6: ...rranty terms and conditions 6 26 Warranty claims will only be accepted if the product is submitted along with proof of purchase in the form of a copy of the invoice or note of delivery The dealer purc...

Page 7: ...G WARNING indicates a hazardous situation that can lead to death or fatal injury DANGER DANGER indicates a maximal hazard due to a situation that can directly cause death or fatal injury 2 2 Purpose P...

Page 8: ...r and malfunctioning from improper storage during long periods of nonuse Reduced product life The medical device should be cleaned serviced and stored in a dry location ac cording to instructions befo...

Page 9: ...n Note We recommend replacing the SONICflex tips every 9 12 months Note Regularly check the wear of the tip with the tip test card Mat no 1 001 6958 The tips may break from continuous stress or damage...

Page 10: ...ing pressure After start ing the intensity can be regulated using the foot switch at maximum drive pressure 3 1 Technical Specifications Drive pressure 2 5 to 4 2 bar 36 to 61 psi Air consumption 26 t...

Page 11: ...ved parallel to the surface of the tooth and not the edge to prevent forming notches in the tooth substance Professionals recommend polishing the tooth surface with the KaVo prophylaxis head using the...

Page 12: ...571 5020 4 First use 4 4 Check the pressure 12 26 Pressure displayed Drive air T R 2 5 to 4 2 bar 36 to 61 psi Return air R L 0 4 bar 6 psi Spray air Sp L max 2 29 bar psi Spray air not required howe...

Page 13: ...NICflex Hold the MULTIflex LUX MULTIflex LED coupling tight and pull the SONICflex forward while twisting slightly 5 3 Inserting the SONICflex tip CAUTION Hazard from a tip which has been incorrectly...

Page 14: ...anti clockwise 5 5 Regulating the spray Rotate the spray ring on the MULTIflex LUX MULTIflex LED coupling in order to regulate the fraction of water The water volume can be regulated by selecting dif...

Page 15: ...Remove the tip from the SONICflex using the torque wrench 6 2 Cleaning CAUTION Malfunctions from cleaning in the ultrasonic unit Defects in the product Only clean manually or in a thermodisinfector 6...

Page 16: ...the following countries Germany Austria Switzerland Italy Spain Portugal France Luxembourg Bel gium Netherlands United Kingdom Denmark Sweden Finland and Norway In other countries interior cleaning c...

Page 17: ...nufactur er of the disinfection agent With KaVo products use only disinfection agents that have been released by KaVo with respect to the compatibility of materials e g WL cid made by ALPRO Immediatel...

Page 18: ...as accessories since they were tested for proper use on our products Note If you are bothered by some oil leaking during the treatment the technical set up allows for switching from servicing with oil...

Page 19: ...Remove the tip Servicing the product See also 2 Instructions for use KaVo QUATTROcare 2104 2104A 6 6 Packaging Note The sterilisation bag must be large enough for the handpiece so that the bag is not...

Page 20: ...dical device has a maximum temperature resistance of up to 138 280 4 F Select a suitable procedure depending on the available autoclave from the following sterilisation processes Autoclave with three...

Page 21: ...1 8 Perio extra long scaler Mat no 0 571 5371 12 Cem Mat no 0 571 5431 Cem attachment Mat no 0 571 7142 16 Retro cylinder left Mat no 0 571 5541 17 Retro cylinder right Mat no 0 571 5581 20 Retro T sh...

Page 22: ...ight Mat no 0 571 7332 Retro plug left Mat no 0 571 5601 Retro plug right Mat no 0 571 5611 58 Bevel mesial Mat no 0 571 7392 59 Bevel distal Mat no 0 571 7382 60 Perio straight Mat no 0 571 7402 61 P...

Page 23: ...Instructions for use SONICflex 2000 N 0 571 5020 7 Tools 6 8 Storage No Tip type Mat no Swing 015 Refill Mat no 1 006 2042 Swing 020 Refill Mat no 1 006 2043 Swing 025 Refill Mat no 1 006 2044 23 26...

Page 24: ...Instructions for use SONICflex 2000 N 0 571 5020 7 Tools 6 8 Storage 24 26...

Page 25: ......

Page 26: ...0 589 0402 kb 20100916 13 en...

Reviews: