Instructions for use INTRAflex LUX 2 Programm 2307 LN - 0.535.6190, 2320 LN - 0.535.6180, 2325 LN -
0.535.6210
7 Reprocessing steps in accordance with ISO 17664 | 7.7 Storage
24 / 28
NOTICE
Contact corrosion due to moisture.
Damage to product.
▶
Immediately remove the product from the steam steriliser after the steril-
isation cycle.
The KaVo medical device has a maximum temperature resistance up to 138
℃
(280.4 °F).
Sterilisation parameters:
Select a suitable procedure (depending on the available autoclave) from the fol-
lowing sterilisation processes:
▪ Steriliser with triple pre-vacuum:
- at least 3 minutes at 135 °C (275 °F)
- Drying time: 16 min.
▪ Steriliser using the gravity method:
- at least 10 minutes at 135°C (275 °F)
- Drying time: 30 min.
▪ Steriliser using the gravity method:
- at least 60 minutes at 121°C (250 °F)
- Drying time: 15 min.
▶
Remove contra-angle handpieces and turbine handpieces from the steam
steriliser immediately after completion of the sterilisation cycle.
▶
Use according to the manufacturer's Instructions for Use.
7.7 Storage
Prepared products must be stored appropriately in a dry, dark, cool room such
that they are protected from germs (as far as possible) and dust.
Note
Comply with the expiry date of the sterilised items.