6.4.2 Cleaning: Automated external cleaning
KaVo recommends thermodisinfectors in accordance with ISO 15883-1
e.g. Miele G 7781/G 7881.
6.4.3 Cleaning: Manual cleaning of the inside
For a medical device with KaVo CLEANspray and KaVo DRYspray.
▶
Cover the medical device with the KaVo CLEANpac bag, and place it
on the corresponding care adapter. Press the spray button three times
for 2 seconds each time. Remove the medical device from the spray
attachment and let the cleaner work for one minute.
▶
Then rinse for 3-5 seconds with KaVo DRYspray.
See also: Instructions for use for the KaVo CLEANspray/KaVo DRYspray.
Preparation methods according to ISO 17664