For program settings as well as cleansers and disinfectants to be used,
please refer to the Instructions for Use of the thermodisinfector (com‐
plying with max. pH value of 10).
▶
In order to prevent negative effects on the medical device, make sure
that the interior and the exterior of the medical device are dry, and then
lubricate immediately with care agents from the KaVo care system.
7.2.3 Cleaning: Manual cleaning - internal
Can only be done with KaVo CLEANspray or KaVo DRYspray.
▶
Cover the medical device with the KaVo CLEANpac bag, and place it
on the corresponding care adapter. Press the spray button three times
for 2 seconds each time. Remove the medical device from the spray
attachment and let the cleaner work for one minute.
Setup methods according to DIN EN ISO 17664
36