Kaufman & Robinson DC3005 Manual Download Page 9

INSPECTION AND INSTALLATION                                                             2-3

_______________________________________________________________________

Copyright © 2011 by Kaufman & Robinson, Inc., 1330 Blue Spruce Drive, Fort Collins, CO 80524
Tel: 970-495-0187, Fax: 970-484-9350, Internet: www.ionsources.com

DISCHARGE POWER SUPPLY MANUAL

MODEL DC3005, MODEL 30010, MODEL 15012

2.4.3 Electrical-connections to ion source and discharge power supply

Located on the back panel of the Power Supply are two sections of three con-

nectors each for connecting the Discharge Power Supply directly to the ion 

source or through a KRI Filament Power Supply.   Each terminal is capable of 

the full output of the Discharge Power Supply.  Multiple outputs are provided 

to allow multiple supplies to share connections.

If the Discharge Power Supply was purchased with an Ion Source and Fila-

ment Power Supply, interconnecting cables are provided.  Information regard-

ing installation of these connecting cables can be found in the “Ion Source 

Filament Cathode Version Controller Manual”. 

If the Discharge Power Supply was purchased as a stand alone unit contact 

Kaufman & Robinson, Inc. for cabling requirements.

2.4.4 Electrical-AC input

The Power Supply has been designed for 85-285 volts AC input voltage, 50-

60 Hz, single-phase with a maximum current of 28 amps at the lowest line 

voltage.  A power supply cable has been included with the Power Supply.  

This cable has a female plug at one end that corresponds to the male socket 

at the rear panel of the Power Supply.  The opposite end of this cable has 

been intentionally left bare so that the cable may be installed to facility re-

quirements.  Electrical connection made from the Discharge Supply Cable to 

the AC voltage within the facility should only be made by quali

fi

 ed personnel. 

Summary of Contents for DC3005

Page 1: ...nson Inc 1330 Blue Spruce Drive Fort Collins Colorado 80524 Tel 970 495 0187 Fax 970 484 9350 Internet www ionsources com Copyright 2011 by Kaufman Robinson Inc All rights reserved No part of this pub...

Page 2: ...TION AND INSTALLATION 2 1 2 2 Inventory 2 1 2 3 Physical Description 2 1 2 3 1 Front Panel 2 1 2 3 2 Rear Panel 2 2 2 4 Installation 2 2 2 4 1 Mounting 2 2 2 4 2 Cooling 2 2 2 4 3 Electrical connectio...

Page 3: ...ww ionsources com DISCHARGE POWER SUPPLY MANUAL MODEL DC3005 MODEL 30010 MODEL 15012 Figures Fig 2 1 Discharge Power Supply front panel 2 4 Fig 2 2 Discharge Power Supply rear panel 2 5 Tables Table 1...

Page 4: ...485 High Speed Serial link for proprietary communications with KRI Controllers Fan on Demand cooling with thermal shutdown protection Front panel selectivity for local remote mode 1 2 Safety Caution O...

Page 5: ...e for the Discharge Power Supply is 85 275 volts ac single phase at 50 60 Hz The maximum input current to the unit is electronically limited to 28 amps RMS Operation at lower line voltages may be limi...

Page 6: ...30 Blue Spruce Drive Fort Collins CO 80524 Tel 970 495 0187 Fax 970 484 9350 Internet www ionsources com DISCHARGE POWER SUPPLY MANUAL MODEL DC3005 MODEL 30010 MODEL 15012 Table 1 1 Maximum Power Volt...

Page 7: ...ny damage has occurred contact Kaufman Robinson Inc in addition to the ship ping company to report any damage see Warranty section 5 Retain packaging materials for shipment of the Power Supply Invento...

Page 8: ...igure 2 2 shows the back panel of the Discharge Power Supply An RS 232 connection is provided on the back panel of this Power Supply for remote operation RS 232 connection at the back panel utilizes a...

Page 9: ...s purchased with an Ion Source and Fila ment Power Supply interconnecting cables are provided Information regard ing installation of these connecting cables can be found in the Ion Source Filament Cat...

Page 10: ..._____________ Copyright 2011 by Kaufman Robinson Inc 1330 Blue Spruce Drive Fort Collins CO 80524 Tel 970 495 0187 Fax 970 484 9350 Internet www ionsources com DISCHARGE POWER SUPPLY MANUAL MODEL DC30...

Page 11: ...Spruce Drive Fort Collins CO 80524 Tel 970 495 0187 Fax 970 484 9350 Internet www ionsources com DISCHARGE POWER SUPPLY MANUAL MODEL DC3005 MODEL 30010 MODEL 15012 163 7 21 1 167 7 21 Input Power RS...

Page 12: ...pply will be in Standby with the Dis charge Amps and Volts Setpoints displayed The Setpoints that are displayed will be the last setpoints used before the supply was switched OFF If the last operation...

Page 13: ...lts Discharge Amps display Displays the discharge current setpoint and run val ues in amps When the yellow setpoint indicator below the display is illuminated the setpoint value is displayed When the...

Page 14: ...of methods Some type of terminal or communications program or programming can be used for controlling this unit Some examples of programs that can be used are Hyper Terminal provided with Microsoft Wi...

Page 15: ...the unit is functioning normally or a number that corresponds to the front panel help codes as described in diagnostics section 4 MEASure CURRent or MEAS CURR Measure current query Returns the DC out...

Page 16: ...e return NOTE If either VX command is zero the unit will go to standby 3 2 4 Operational sequence Using a terminal program configured for the communications settings previ ously mentioned input the fo...

Page 17: ...t VA0 then press the enter key The out put will drop to zero and the Standby LED will be illuminated Output Current Monitoring of the output current can be done using the following command Input RD0 t...

Page 18: ...Supply for connection to the analog interface The pin descriptions for the interface are given below in table 1 There is a button on the front panel that can be used to switch from remote to local co...

Page 19: ...on nected to common The shutdown is deactivated when 5 volts is applied or no contact is made Remote shutdown is indicated on the front panel when activated 15 Auxiliary 5 volts DC 200 mA max Also ava...

Page 20: ...ruce Drive Fort Collins CO 80524 Tel 970 495 0187 Fax 970 484 9350 Internet www ionsources com DISCHARGE POWER SUPPLY MANUAL MODEL DC3005 MODEL 30010 MODEL 15012 Table 3 4 Interlock connections Pin Nu...

Page 21: ...gas bottle to the ion source Power must be removed from the Power Supplies prior to performing maintenance on the Ion Source 4 2 Diagnostic Table The following table may be used to assist in determin...

Page 22: ...e PC Shows up only in re mote RS 232 operation Display says HLP 8 for 5 seconds Informational display only No action taken on unit operation Unit recovers from error condi tion if the problem goes awa...

Page 23: ...ts KRI s obligation under this Limited Warranty is at KRI s option to repair or replace any defective equipment or parts of the equipment without charge to the purchas er which are returned shipping p...

Page 24: ...oblem encountered or the item to be returned Product Serial number Detailed description of problem Date of purchase Name and address of company Contact person If return to KRI is required you will be...

Reviews: