• For separate measurement of flow and temperature select a

flow unit as

Middle Unit in the Setup Wizard. Complete the

wizard and then go to

Main Menu - In/Output.

• In the

In/Output menu assign the temperature input to

channel 1, and select

Inlet or Outlet from the menu.

• In the

Display menu set the Top Line or Bottom Line to either

Tin or Tout depending on the previous selection.

• In the

Datalogger menu select either Tin or Tout from the list

of variables depending on the previous selection.

• When measuring, the flowmeter will be logging the PT100 input

and can be viewed by pressing

1/NEXT on the keypad.

• This data can also be assigned to a process output in the

In/Output menu following the same process.

Process In/Outputs: Setup

• Configured in/outputs can be set up in

Main Menu - In/Output,

depending on which in/outputs are available. The in/output lis-

ted at the top of the list on

Slot 1 inside the connection compart-

ment, the in/output listed second from top on

Slot 2 and so on.

• After selecting an in/output, it can be assigned to a measurement

channel. The structure of the further steps depends on the indivi-

dual type of in/output.

Process In/Outputs: Connection

• In/outputs can be connected to devices via screw terminals.

These can be found inside the connection compartment below

the control panel.

• The illustration on the left gives an overview of the terminal

assignment (Pic. 6).

• Remove the screws at the left and right to gain access to the

compartment.

Heat Quantity Measurement (where specified)

• The KATflow 150 is capable of measuring heat flow and heat quantity.
• To measure heat flow, select a heat flow unit (W, kW, MW) as

Middle

Unit when going through the Setup Wizard of the selected measure-

ment channel.

• To measure heat quantity, select a heat quantity unit (J, kJ, kWh).

• On selection of one of these units, the flowmeter requests the entry

of the specific heat capacity of the medium in [J/(g.K)]. Enter the

value of the fluid and confirm with

ENTER. Complete the remaining

steps ofthe

Setup Wizard.

• The temperatures at the inlet and outlet of the monitored system are

used to determine heat flow and heat quantity.

• Connect the supplied 4-wire PT100 temperature sensors to the flow-

meter as per the illustration below (Pic. 6).

6. Input/output compartment, inside terminals

• After setting up the flow measurement channel go to

Main Menu -

In/Output, in order to assign the PT100 inputs to the channel.

• Use the cursor keys to select

PT100 4 WIRE and confirm with ENTER.

• On the next screen, use the cursor keys to select the flow channel to

which the temperature input is applied. Choose

Channel 1 or 2.

Selecting

OFF will disable the input. Confirm with ENTER.

• On the following screen choose

PT100 to use the temperature meas-

ured on the pipe. To enter a fixed temperature value, select

User and

enter the value.

• Define whether the PT100 sensor measures the inlet or outlet tempera-

ture. Use the cursor keys to select as appropriate and confirm with

ENTER.

• On the next screen a temperature offset can be applied. After confirm-

ing again with

ENTER the meter returns to the Main Menu.

• Once the setup of the first temperature input has been completed, re-

peat these steps for the configuration ofthe second input.

Temperature and Flow Measurement

3/3

© Copyright Katronic Technologies Ltd. 2015 | Issue: QS_KF150_V42EN_1510 | All rights reserved.

KATflow 150

Quick Start Manual