K-2oem ~ ALT-AIRE+ QUICK REFERENCE GUIDE 

 

K-2oem ~ ALT-AIRE+™ 

ALTERNATING THERAPY WITH ON-DEMAND LOW AIR LOSS SYSTEM

 

 

THIS DOCUMENT IS FOR QUICK REFERENCE ONLY.  PLEASE READ THE MANUAL BEFORE OPERATING THE K-2oem SYSTEM.

 

Note:

 Before using the KAP MEDICAL Mattress or Overlay System, please remove current mattress,  

          

mattress replacement systems or overlay systems from the bed. 

 

 

Mattress / Overlay set up. 

 Make sure the mattress/overlay fits snugly on the bed frame without any gaps on the side. 

 

Overlay System: Place the K-2oem overlay directly on an existing 3” to 5” foam mattress. Insert head and 
foot end elastic straps around the foam mattress. Loop each long side strap around the foam mattress and 
fasten it securely to the short strap on the other side using the buckle. Hang the K-2oem control unit on 
the footboard, make sure that the mattress CPR valve is in closed position and hose assembly is 
connected securely to the control unit. 

 

Mattress Replacement: Unroll the K-2oem mattress replacement and place it directly on the bed frame 
and attach it firmly with the straps provided.  Hang the K-2oem control unit on the footboard, make sure 
that the mattress CPR valve is in closed position, bolster and/or air pad plugs are connected properly and 
check that the mattress hose assembly is connected securely to the control unit.   

 

Plug the hospital grade power cord provided with the unit into a three-pronged hospital socket.  The amber 
light will illuminate.  Press POWER key 

(SB)

 to green light position and then press Select key 

(S) 

until MAX 

Inflate LED 

(W) 

lights up

.

  The compressor will come on and quickly inflate the mattress. 

 

Using the SOFT / FIRM keys 

(K)

 set patient comfort pressure settings.  7 segment display will show patient 

comfort settings from 0 (lowest pressure) to 9 (highest pressure)

.

 

 

For ALTERNATING PRESSURE 

(AP) 

therapy, press Mode 

(M)

 key until the A/P LED 

(AP)

 lights up. Press 

Time key

 (TT)

 to select the desired cycle time for AP therapy.   

 

To set

 

STATIC 

(T)

 mode, press Mode key

 (M)

 until “STATIC” LED 

(T)

 lights up. 

 

To activate Low Air Loss 

(Y)

 press the Select key 

(S)

 until LAL LED 

(Y) 

lights up. 

 

To set ALARM SILENCE, press the ALARM SILENCE key 

(AS) 

until the indicator light goes on. 

 

To set the LOCK OUT Function, press the LOCK OUT key 

(LO)

 until indicator light goes on.  

 

K-2oem is equipped with a Fowler key 

(U) 

(press Mode key 

(M)

 until Fowler LED 

(U) 

lights up) which when 

pressed in the patient’s fowler position of 35 degrees or greater will inflate the mattress to higher pressures 
to eliminate patient bottoming.   

 

For 

CPR

 or quick deflation, disconnect mattress hose connector from the control unit by simply pressing the 

latches on the quick disconnect connectors and pulling the connector coupling from the control unit in a 
straight outward direction. Rotate the CPR valve on the mattress / overlay to the OPEN position for quick 
deflation (CPR valve not present on FOAM AIRE mattresses). 

 

 

 For more information, please call authorized service center 

  

 

Summary of Contents for K-1 Series

Page 1: ...SYSTEM K 2 ELITE KARE ZONE DIGITAL ALTERNATING PRESSURE WITH ON DEMAND LOW AIR LOSS SYSTEM K 2oem ALT AIRE ALTERNATING PRESSURE WITH ON DEMAND LOW AIR LOSS SYSTEM US INTERNATIONAL PATENTS PENDING Ple...

Page 2: ...ice 1 866 KAP MED 1 866 527 6331 Technical Support Service 1 951 340 4366 Email sales kapmedical com service kapmedical com www kapmedical com FDA Registered Company California FDA Registered Company...

Page 3: ...ail Entrapment Risk Notification 9 Unit Overview 11 Explanation of Symbols Used 13 Mattress Overview 15 Unpacking the System 16 Mattress System Set Up 17 Operating Instructions 18 CPR Function 23 Pati...

Page 4: ...pport combustion Failure to observe this warning can result in severe fire property damage and cause injury or death Entrapment may occur Patient entrapment with bed side rails and mattress may cause...

Page 5: ...harmful interference received including interference that may cause undesired operation Changes or modifications not expressly approved by the party responsible for compliance could void the user s a...

Page 6: ...2 LPM Support Surface Inflation Time 5 15 Minutes Patient Comfort Pressures Alternating Pressure Soft Pressure 6 5 mmHg K 1 K 2 Firm Pressure 32 5 mmHg K 1 31 5 mmHg K 2 of high AP pressure setting A...

Page 7: ...n height replacement mattresses available along with 3 or 5 mattress overlays Packaging 1 piece per box Patented Low Air Loss top sheet on K 2 and k 2oem models only Optional mattress with Kevlar line...

Page 8: ...California117 Optional California TB 106 TB 129 16 CFR 1632 16 CFR 1633 BS 6607 CRIB 5 BS 597 1 BS 597 2 compliant mattresses lined with Kevlar fire barrier available Kevlar lining based on flammabil...

Page 9: ...control unit is not blocked and kept away from any heat sources or radiators during the operation of the unit Care should be taken such that the power cord of the control unit is not pinched or has an...

Page 10: ...patient is unique Only the patient s medical care provider is familiar with the patient s unique medical condition and needs Only the patient s medical care provider and or the dealer from whom you o...

Page 11: ...ces TSS which support the patient on an air mattress or specialized foam layer may present an increased risk of entrapment for some patients The benefit of TSS product use must be weighed against the...

Page 12: ...Thermoplastic 94V 0 fire retardant enclosure houses a compressor main PCB display panel a short circuit over voltage protection with single dual fuse a quick connection and disconnection CPR deflatio...

Page 13: ...s Owner s Manual Page 12 Refer to the Explanation of Symbols Used chart on page 13 for a cross reference of Display Key descriptions Unit Display Labels K 1 DERMA KARE K 1oem ALT AIRE K 2 KARE ZONE K...

Page 14: ...STATIC T or Selects Static Therapy mode LOW AIR LOSS Y or or Activates Low Air Loss mode ALTERNATING PRESSURE AP or or Selects A P mode AP TIME TT or Selects AP cycle time FOWLER U or Fowler Mode The...

Page 15: ...luded with each model SYMBOL EXPLANATION No Sharp Objects Low Heat Setting Do Not Dry Clean Do Not Bleach No Open Flames Normal Cycle Do Not Iron Latex Free Bolstered Mattress Even Bolstered Mattress...

Page 16: ...top sheet has a special multi chambered air distribution layers which administers low air loss relief directly underneath the patient Top layer made out of urethane coated 70 Denier nylon or Dartex o...

Page 17: ...of moving bed components before placing a patient on the bed and that the mattress system is fully inflated Operate all bed frame motorized functions through their full range of motion to be certain...

Page 18: ...mattress system care should be taken such that the mattress is placed directly on the bed frame Figure 1 B Page 16 Unroll the Replacement Mattress and place it on the bed frame BF There are ten nylon...

Page 19: ...m initialization routine for a few seconds Once the routine is completed the display D will read KAP MEDICAL STANDBY K 1oem K 2oem During initial power up when power cord Q is plugged into the power s...

Page 20: ...atient s body There should be a minimum of 3 to 4 finger width clearance between the bottom of the patient and the safety foam base on an overlay there is no safety foam base Repeat this procedure unt...

Page 21: ...even numbered air cushions deflate to the set percentage of the patient comfort high pressure in the first half of the AP cycle and vice versa for the second half of the cycle and continue back and f...

Page 22: ...is factory programmed to a 35 or 45 angle depending on the customer s requirements It uses a 3 6V Lithium battery that can be replaced using KAP Medical part number 100622 S When the battery is first...

Page 23: ...Cushion K 140 OPTIONAL SAC Optional foot support air cushion can be used to support patient s feet and to provide zero interface pressures in the heel area FOAM AIRE MATTRESS AS A NON POWERED MATTRESS...

Page 24: ...mattress for a CPR procedure pull the hose V from the control unit connector press connector latches then pull connector If the mattress is equipped with bolsters and or air pad unplug the correspond...

Page 25: ...unit flange connector PLEASE NOTE Before using the mattress please make sure that the mattress deflate connector is re connected back into the mattress deflate valve PATIENT TRANSPORTATION 1 Air Matt...

Page 26: ...ray the exterior of the top and the bottom enclosures power cord and the cord plug with the prepared disinfectant detergent solution Using a damp cloth wipe down the sprayed surface cleanly Note Do no...

Page 27: ...ote The Foam Aire mattresses have a Kevlar fire retardant sleeve sock barrier inside the cover Care should be taken if removing the cover for cleaning Before attempting to clean the mattress remove th...

Page 28: ...o power source 3 Check power source has power and turn it ON 4 Wait till the power source has power 5 Send control unit back to factory for repair C Control Unit not responding 1 Unit locks up 1 Unplu...

Page 29: ...ressure setting or 7 75 mmHg of high comfort pressure setting Use the Soft Firm keys to toggle between AP Low modes To end routine and accept setting press Power key Routine will auto time out K 2 ELI...

Page 30: ...ICAL be liable for any loss direct indirect incidental or special damages arising out of or in connection with the use of this product The control unit warranty does not cover normal maintenance such...

Page 31: ...K 1 K 2 Series Owner s Manual Page 30 NOTES Unit S N __________________________________ Mattress S N _______________________________ Auto Fowler S N ____________________________...

Page 32: ...nnected securely to the control unit Plug the hospital grade power cord provided with the unit into a three pronged hospital socket The display will read KAP MEDICAL STAND BY Press POWER key SB to gre...

Page 33: ...tress hose assembly is connected securely to the control unit Plug the hospital grade power cord provided with the unit into a three pronged hospital socket The amber light will illuminate Press POWER...

Page 34: ...o set the patients height and weight Patient comfort settings will be automatic when done with settings Adjust the SOFT FIRM knob K to bypass automatic patient comfort pressure settings Actual mattres...

Page 35: ...e power cord provided with the unit into a three pronged hospital socket The amber light will illuminate Press POWER key SB to green light position and then press Select key S until MAX Inflate LED W...

Reviews: