18
5.4 CONTINUOUS Mode
It is a mode not to set the sample time, and nor to begin, and to end the measurement with the START/STOP key.
Particle size(µm)
、
data store(STR)
、
Warning(BEEP)
、
Printer(PR) can be set.
Display screen
Operation
key
Operation explanation
POWER
Push POWER key to input the power supply.
PREV
Push PREV key twice to proceed MODE screen
△
▽
set
Select 3. CONT
△
▽
set
■
Setup the particle size, requirement of data
storage , alarm , printout.
Use
△
▽
key to change the setting
Condition, then push SET key.
After the input is done, change NO to OK
and push SET key.
WAIT Mode
WAIT sign is shown for the stabilization of internal
pump
READY Mode
WAIT sign turns to READY in 10 seconds and
measurement can be started.
Push START/STOP key to start the measurement
START
/
STOP
The display shows the real-time data.
Remaining time of each measurements
⑦:
Measurement time
(
Count up
)
★
“
01h00m
”
and the display change into the
following of 59:59 (It is 59second of 59 minutes.)
START
/
STOP
START/STOP key is pushed, and the
measurement is ended.
⑧:
Measurement time
★
The data is printed after the sampling time is
Over if you chose printout. (refer to 6.4)
CONT
0.3
/
0.5µm
STR:Y BEEP:N PR:N
OK
CONT 003BPM 15:25
WAIT 0.3 0.00E+0
/
cf
01
/
06 0.5 0.00E+0
/
cf
CONT 003BPM 15:25
READY 0.3 0.00E+0
/
cf
01
/
06 0.5 0.00E+0
/
cf
CONT 004BPM 15:25
00:01 0.3 0.00E+0
/
cf
0.5 0.00E+0
/
cf
⑦
→
CONT
004BPM 15:58
STOP 0.3 0.00E+0
/
cf
32:48 0.5 0.00E+0
/
cf
The particle size can be switched with the
△
▽
key on the measurement screen.
UNIT: [CNT] [/m
3
] [/cf]
<MODE>
1. REPEAT 4.CALC
2. SINGLE 5.REMOTE
3. CONT
⑧
→