Page 1: ...1 User Guide 200 netzerotools com Kanomax S200 Handheld Multi Gas Monitor netzerotools com...

Page 2: ...concentrations above the recommended maximum range Do not place the sensor head in a gas stream the monitor is designed to measure ambient levels of ozone and uses active sampling to achieve this 1 Se...

Page 3: ...4 5 Zero Calibration 11 4 6 Changing Unit of Measurement 12 4 7 Max Min Average Measurement Cycle 12 4 8 Alarm Mute Perc only 13 4 9 Alarm Perc only 13 4 10 Alarm Level Perc only 14 5 0 Troubleshootin...

Page 4: ...y of the components are missing Replacements Replacement batteries sensor heads and other accessories can be purchased separately Note All sensors deteriorate over time It will be necessary to replace...

Page 5: ...user the ability to replace faulty sensor heads without having to replace the monitor The monitor can be mains and or battery pow ered and either permanently mounted or used as a portable unit 2 The...

Page 6: ...be installed and charged prior to use 3 1 Battery Pack Remove the battery cover from the monitor Insert the battery power connector into the monitor power socket in the back of the monitor Lay the ba...

Page 7: ...nitor Do not disassemble or mutilate the batteries as this may cause burns Do not incinerate or heat as this may cause burns the batter ies may burst or cause the release of toxic materials Do not sho...

Page 8: ...the base Remember to switch off the monitor before removing the sensor head Unplugging the main power will reset the above charging condi tions The batteries can be charged and discharged many times...

Page 9: ...rming up or sensor failure in older models to disappear Should the message remain after this time a genuine sensor failure may have occurred See the trouble shooting guide Since O3 is an oxidising gas...

Page 10: ...the display will change to indicate the type of sensor head The monitor will warm up as described in Warm up in section 3 2 Once warm up is complete the main display will appear for example To turn t...

Page 11: ...full power and the fan remains on 4 2 Operating under Battery power A battery indicator will appear at the bottom of the display when the monitor is operating under battery power only The indicator sh...

Page 12: ...C 68 F 3 5 F Relative Humidity of 50 5 An incorrect ZERO CAL does not cause permanent damage The process should be repeated under the above conditions NOTE Carbon Dioxide sensors MUST be calibrated in...

Page 13: ...s the Min Max Av routine Press to scroll to EXIT Press to Exit out of Menu Setup The following display will appear Repeat the above sequence to stop the Min Max Av routine Note The MIN MAX and AVE rea...

Page 14: ...RM MUTE ON Press to confirm your choice and you will return to the main menu If you select Alarm Mute On and exit the menu the symbol will be displayed on the gas concentration screen Press to scroll...

Page 15: ...een will appear Press to toggle between 50PPM 100PPM 150PPM and 35PPM Press to confirm your choice and you will return to the main menu Whatever alarm level you have chosen will be displayed on the bo...

Page 16: ...under zero gas conditions Background gas level higher than normal Interferent gas present Sensor damaged Move sensor to clean air and recheck baseline Move sensor to clean air and recheck baseline Re...

Page 17: ...erator Move sensor to clean air and check reading upon exposure to known gas concentration Replace refurbish sensor Reading unstable Power supply unsta ble Power supply current rating incorrect Local...

Page 18: ...s control None Series 200 Monitor Base External signal functions None Analogue output None Power supply 12 VDC 800 mA Rechargeable battery pack 9 6V 2100mA hr Ni MH Portable Yes Permanently fixable Sc...

Page 19: ...18 Depressions in the battery cover demarcating the screw hole posi tions for fixing the unit to the wall 7 0 Drawings netzerotools com netzerotools com...

Page 20: ...mage the electronic componentry Do not attempt to open Non expert handling of the device may cause damage Do not drop knock or shake as this could lead to internal damage Do not use harsh chemicals cl...

Page 21: ...hen connecting to any other device read the appropriate user guide for detailed safety instructions Do not connect in compatible products HAZARDOUS ENVIRONMENTS Ozone Sensors Only Do not use the Ozone...

Page 22: ...protection against harmful interference in a residential installation This equipment gener ates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruc...

Page 23: ...subsequent period as stated in the following table This warranty is expressly limited to the original owner who purchases the equip ment directly from Aeroqual or from an authorized Aeroqual dealer W...

Page 24: ...d by use of non Aeroqual original accessories This warranty becomes void if a non Aeroqual approved AC DC adaptor or battery is used Tampering with any part of the product will void the warranty Damag...

Page 25: ...cribed in this document without prior notice Under no circumstances shall Aeroqual be responsible for any loss of data or income or any special incidental consequential or indirect dam ages howsoever...

Page 26: ...ler has been designed to meas ure the ambient concentration of ozone The controller must not be placed directly in an ozone stream For indoor local area monitoring attach the controller to an inert su...

Page 27: ...d in the past week There are trace levels of gases present in the air to which the sensor is cross sensitive In many cases the non zero base line reading occurs in an environment which is supposed to...

Page 28: ...ne standard 0 25 ppm Ozone O3 0 0 150 ppm 0 0 5 ppm 0 10 ppm Leak detector Perchloroethylene C2Cl4 0 200 ppm Sulphur dioxide SO2 0 10 ppm 0 100 ppm Volatile Organic Compounds VOC Isobutylene standard...

Reviews: