Manual KanMed Operatherm 

 

Page 

5

 

 

 

 

3 TECHNICAL 

SPECIFICATIONS 

 

Heating Pad Standard  

Art. no.:OP-200-002

 

 

Voltage 

 

30 Volt DC 

 Power 

  

200 

Watt 

 

Dimension 

 

L 1040 mm  W 450mm  H 2 mm 

 Weight 

  

kg 

 

Length of cable 

3,5 meter approx. ( Other lengths may be ordered ) 

 

Water protection 

Watertight, except connector 

 

Surface, electrical properties 

Not electrical conductive 

 

Heating Pad Small 

       

 

 

Art. no.: OP-200-012 

 

Voltage              

 

 

30 Volt DC 

           Power                

 

 

100 Watt 

      

Dimension         

 

 

L 630 mm  W 430 mm  2 mm 

 Weight 

    1,5 

kg 

 

Length of cable        

 

3.5 metre approx. (Other lengths may be ordered) 

 

Water protection 

Watertight, except connector 

 

Surface, electrical properties.  

Not electrical conductive 

 

Heating Pad 150cm  

Art. no.: OP-200-221

 

 

Voltage 

 

30 Volt DC 

 Power 

  

200 

Watt 

 

Dimension 

 

L 1500 mm  W 450 mm  H 2 mm 

 Weight 

 2,8 

kg 

 

Length of cable 

3,5 metre approx 

 

Water protection 

Watertight, except connector 

 

Surface, electrical properties 

Not electrical conductive 

 

Control Unit 

  

 

Voltage              

 

 

220 -240 V AC ( 100 and110-120 V AC is also available ) 

    

Frequency            

 

 

50/60 Hz 

   

Power                

 

 

285 VA (max) 

    

Temperature range    

 

32°C and 35-39°C in steps of 0.5°C 

 

 Accuracy 

   0.5°C 

 

 Indicators 

   LED 

 

 Resolution 

   0.5°C 

 Over-temperature 

alarm 

level 40,25°C 

±

  0.75°C 

 

Dimensions       

 

 

L  255 mm approx. W  150 mm approx H  120 mm approx 

 

Weight               

 

 

5 Kg approx. 

     

Fuses                

 

 

220 - 240 Volts  T 4A,  110 –120 Volts  T 8A 

 Mode 

of 

use 

   Continuous 

 

Heating up time

 

 

 

Elements inside the Heating Pad 

 

about 8-10 minutes at room temperature 

 

Heating Pad surface

   

 

 about 25 minutes when tested under conditions of  
ADEQUATE HEAT DISCHARGE 

(IEC601-2-35)

 :   

 

Environment  

 

 

Operating temperature 

 

15°C - 40°C 

 

Shipping and storage  

 

-20°C - + 60°C 

 

Relative humidity 

 

 

10 - 85 % non condensing 

 

 

 

 

 
Safety Standards 

 

   EN60601-1, 

EN60601-1-2, 

IEC60601-2-35 

 

Protection type  

 

 

BF, Defibrillator safe 

 

Protection class  

 

 

Class I

  

 

MDD classification   

 

Class IIB 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0413

 

 

 
NOTE: Specifications are subject to changes at any time. 

 

Summary of Contents for Operatherm 202

Page 1: ...NANCE 12 NOTE Patient warming equipment can harm the patient if not used correctly Never use a unit that do not perform the Self Test according to this manual Make sure that the performance of the uni...

Page 2: ...t has been examined by a qualified technician The Operatherm 202 is a very safe equipment However certain combinations of heat pressure long term surgery and disinfectant fluids can cause skin damage...

Page 3: ...safety 4 The Heating Pad cable may not be altered in length or repaired in any other way 5 See Users Manual section 10 concerning cleaning and maintenance of Heating Pads and Control Units 6 Sharp obj...

Page 4: ...nt 2 8 TEMPERATURE SETTING Under normal conditions and with KanMed Gel Pads under the patient the maximum temperature 39 C should be selected During long interventions one must consider a lower temper...

Page 5: ...Watertight except connector Surface electrical properties Not electrical conductive Control Unit Voltage 220 240 V AC 100 and110 120 V AC is also available Frequency 50 60 Hz Power 285 VA max Temperat...

Page 6: ...is switched on it will perform a self test of all functions including the internal safety circuit This is how the self test runs When pressing the ON OFF button all front panel indicators and the beep...

Page 7: ...ong pad or of the 4 elements in the short pad should become warmer than 40 25 C The visual and acoustic alarm will be activated and the over heated element s will be switched off for 90 sec The acoust...

Page 8: ...ts disconnected during normal use the pad fault indicator and the 32oC LED will be lit The beeper will also sound Switch the unit off and reconnect the Heating Pad Restart the unit and observe a corre...

Page 9: ...a technician 5 3 SELECTING TEMPERATURE In case you wish to select another temperature than 37 C then do as described in chapter 4 4 1 It will take about 8 15 minutes for the elements in the pad to re...

Page 10: ...1 2 MAINS POWER FAILURE ALARM AND RESTARTING Connect the Heating Pad to the Control Unit and the Control Unit to the mains Switch on the unit and wait for the self test to be completed Disconnect the...

Page 11: ...tact your local dealer 8 EMC COMPATIBILITY STATEMENT Please see www kanmed se section downloads Operatherm 9 WARRANTY STATEMENT KanMed warrants the purchaser that the Operatherm Control Unit and Heati...

Page 12: ...ied out by the user operator Self test at Power ON Every time the equipment is switched on Visual inspection of pad cable and connector Before every use Checks carried out by technical personnel Alarm...

Reviews: