2.3
Flow data
Nom. flow q
p
Nom. diameter
Meter factor
1)
Dynamic range
Flow@125 Hz
2)
D
p@q
p
Min. cut off
[m³/h]
[mm]
[p/l]
q
p
:q
i
q
s
:q
p
[m³/h]
[bar]
[l/h]
150
DN150
1
100:1
2:1
450
0.02
300
250
DN150
0.6
100:1
2:1
750
0.055
500
400
DN150
0.4
100:1
2:1
1125
0.04
800
400
DN200
0.4
100:1
2:1
1125
0.01
800
400
DN250
0.4
100:1
2:1
1125
0.01
800
600
DN200
0.25
100:1
2:1
1800
0.022
1200
600
DN250
0.25
100:1
2:1
1800
0.022
1200
1000
DN250
0.15
100:1
2:1
3000
0.015
2000
1000
DN300
0.15
100:1
2:1
3000
0.015
2000
1)
Standard meter factor. Appears from ULTRAFLOW
®
label. For other meter factors see paragraph 4.
2)
Saturation flow 125 Hz. Max. pulse frequency is maintained at higher flow.
Table 1. Flow data.
2.4
Materials
Wetted parts
Housing
Stainless steel, W.no. 1.4307
Transducer holder Stainless steel, W.no. 1.4308
Transducer
Titanium
Gaskets
Fibre
Electronics box
Base, cover
Thermoplastic, PC 10 % GF
Fitting hardware,
Thermoplastic, PPS 40 % GF
distance piece for
electronics box
Signal cable (optional for separate ULTRAFLOW® 54)
Silicone cable (3 x 0.5 mm
2
)
Power supply cable 24/230 VAC (optional when selecting supply module)
Cable with PVC-mantle (2 x 0.75 mm²)
Housing, Cable Extender Box
Base, cover
Thermoplastic, ABS