User Manual
PKK-50
24
Press
To choose between programs press on the number (below display) and press .
When program was stopped by pressing stop button, you get notification on display
If you would like to continue program from the point where it was interupted press
Or if you would like to start from beginning press
After few seconds
Meaning:
STEP Nr: 1: 30:00
First step in the process. Time set to 30 minutes [when temperature
and Rh reaches the tolerance area the time count down starts. See
special settings]
Tc=38.3°C / 40°C
Current chamber temperature is 38.3°C, set chamber temp. 40°C
Rh= 62.2% / 75 %
Current chamber humidity is 62.2%, set humidity is 75 %.
CO2= 0.5% / 20 %
Current chamber CO2 concentration is 0.5%, set is 20 %.
1
SELECT PROGRAM
TO START
(Program name)
SELECTED PROGRAM: 1
STEP Nr: 1 : 30:00
Tc= 38.3°C / 40°C
Rh = 62.2 % / 75%
CO2= 0.5% / 20%
PROGRAM INTERUPTED!
PRESS BUTTON
-> TO CONTINUE PROGR.
<- FROM BEGINNING
WARNING:
This kind of display indicates automatic running of selected program. The machine will
automatically turn off at the end of all temperature/rh steps!