4.2.3 Unlock screen 

 
Press Unlock button to see this screen: 

       
       
       

       
       

       
       
       

       
       

       
       

       
       
 

Enter unlock code 0207 and confirm with OK button to enable edit PID settings and temperature corrections. 
 

To disable edit PID setting and temperature corrections press Lock button. 
If you forget to press Lock button after 2 minutes it locks automatic.  
       

       

4.2.4 Graph screen 

 
Press Graph button to see this screen: 

       

       
       

       
       
 

 
 

 
 

 
 
 

 
 

 
 
 

 

 
 

 
 
 

 

Enter unlock code 0207 

Press to Lock 

Press OK to confirm and unlock   

Enter full screen button 

 

Back to Main menu button 

 

Slider to move graph 

 

Slider to zoom graph in y 
direction 

 

Slider to move graph 

 

Slider to zoom graph in x 

direction 

 

Summary of Contents for OB-22/2 LT

Page 1: ...User Manual Calibration Bath OB 22 2 LT ...

Page 2: ...ofits loss of income loss of business opportunities loss of use and other related exposures caused by incorrect use of product Warranty is valid for 13 months from date of shipping and includes spare parts and labor Type OB 22 2 LT Serial number Date of shipping Signature stamp Manufacturer Kambic d o o Metliska cesta 16 8333 Semic Slovenia Tel 386 7 35 65 220 Fax 386 7 35 65 232 Email info kambic...

Page 3: ... 18 4 2 4 Graph Screen 18 4 2 5 Main Settings 19 4 2 6 Pid Parameter Settings 19 4 2 7 Special Settings 20 4 2 8 Compressor Cooling Settings 20 4 2 9 Temperature Corrections 21 4 2 10 Stop Screen 22 4 2 11 Back Panel 22 5 SERVICE AND MAINTENANCE 23 5 1 Back Panel 24 6 REQUIRES SPACE AND CONECTIONS 25 7 SAFE USE AND SAFETY ELEMENTS 26 8 DISMANTLING AND DISPOSAL 27 9 TROUBLESHOOTING 28 11 RS 232 COM...

Page 4: ...not sure about safe use of the product in your specific case do not hesitate to contact your dealer or us via info kambic com for further information or opinions In no event shall Kambic d o o be held liable for any damages direct or incidental related to the use of this product and manual Product can be equipped with variety of accessories and options This warning list might include elements that...

Page 5: ...t operate the equipment in wet and condensing environment STABILITY HAZARD Do not load or affect product stability with heavy loads or climbing the product TRANSPORT HAZARD Transport the chamber only in original packing Always secure the chamber stability during reposition Never use doors or handles or accessories for holding point during reposition FREEZING HAZARD Do not store or transport produc...

Page 6: ...here Verwendung des Gerätes in Ihrem spezifischen Fall unsicher sind zögern Sie nicht Ihren Händler oder uns über info kambic com zu weiteren Informationen oder Stellungnahmen zu kontaktieren Keinesfalls kann Kambic d o o für Schäden jeglicher Art durch Verwendung dieses Produktes oder Handbuchs direkt oder indirekt verantwortlich gemacht werden Das Produkt kann mit verschiedenen Zubehörteilen und...

Page 7: ...densierender Umgebung betreiben GEFAHR DER SCHWERPUNKTVERLAGERUNG Keine schweren Lasten laden oder Stabilität durch Zuladung oder Besteigen des Produktes beeinflussen GEFAHR DURCH TRANSPORT Transportieren Sie die Kammer nur in der Originalverpackung Bei Positionsveränderung stets die Kammerstabilität absichern Niemals Türen Griffe oder Zubehörteile zum Tragen des Gerätes nutzen FROSTGEFAHR Nicht e...

Page 8: ... des doutes concernant une utilisation sûre dans votre cas précis n hésitez pas à contacter votre vendeur ou nous mêmes via info kambic com pour plus d informations ou de conseils En aucun cas Kambic d o o ne pourra être tenu responsable pour quelques dommages directs ou accidentels que ce soient liés à l utilisation de ce produit ou de ce manuel Le produit peut être équipé de nombreux accessoires...

Page 9: ...environnement humide et présentant de la condensation RISQUE D INSTABILITÉ N affectez pas la stabilité du produit avec de lourdes charges ou en montant sur celui ci RISQUES LIÉS AU TRANSPORT Ne transportez la chambre que dans son emballage d origine Assurez toujours la stabilité de la chambre pendant son déplacement N utilisez jamais de portes de poignées ou d accessoires comme point d appui penda...

Page 10: ...o está clara o tiene dudas en cuanto al uso seguro del producto en su caso particular contacte con su distribuidor o escríbanos a info kambic com para obtener información adicional u opinión Sociedad limitada Kambic d o o en ningún caso será responsable de ningún tipo de daños directos o incidentales resultantes de la utilización de este producto y manual El producto puede ser provisto de una vari...

Page 11: ...S No manipule el equipo en un ambiente húmedo y de condensación RIESGO DE INESTABILIDAD No aplique carga ni desestabilice el producto con cargas pesadas No subir al producto RIESGOS AL TRANSPORTAR Transporte la cámara sólo en su empaque original Proporcione siempre estabilidad a la cámara al reposicionarla Nunca la agarre por las puertas asideros o accesorios durante el reposicionamiento RIESGO DE...

Page 12: ...нформация для вас является неясной или вы не уверены в безопасном использовании изделия в вашем конкретном случае рекомендуем обращаться к вашему дилеру или к нам через info kambic com за дополнительной информацией или мнением Ни при каких обстоятельствах Kambic d o o не несет ответственности за любые повреждения прямo или косвенно связанные с использованием данного изделия и руководства Изделие м...

Page 13: ...йте оборудование во влажной и конденсационной среде ОПАСНОСТЬ НЕСТАБИЛЬНОСТИ Не перегружайте и не влияйте на стабильность изделия тяжелыми нагрузками или подъемом на изделие ТРАНСПОРТНАЯ БЕЗОПАСНОСТЬ Транспортируйте камеру только в оригинальной упаковке Всегда сохраняйте стабильность камеры во время перемещения Никогда не используйте двери или ручки или аксессуары для Удерживания во время перемеще...

Page 14: ... working volume Of course the construction is not enough that is why we are using high capacity microprocessor controllers and different levels of heating and cooling A touch screen display is used to set the individual temperatures All controllers are also equipped with communication modules so the appliance can be operated and controlled via a PC The appliance is completely made of stainless ste...

Page 15: ... 130 C Temperature Stability Better than 0 005 C Temperature homogeneity Better than 0 007 C Display resolution 0 001 C Set up resolution 0 001 C Control MPC PID Power supply 10 230 V 50 Hz Wattage 3200 W Communication port RS 232 Built in Cooling Air cooled compressor refrigeration system Construction materials SS vessel exterior housing cover Temperature probe PT 100 ...

Page 16: ...230V 10 50Hz Step 3 Put the main switch in position 1 on the back of the device Step 4 Press On off switch Step 5 Select the desired temperature and press START button Step 6 Check the fluid level and add if necessary Step 7 Press STOP to stop the cycle 4 2 Control panel with touch display START button On off switch STOP button LCD touch display Power LED Operating LED ...

Page 17: ...lock button Graph button Heating power Firmware version date and time Fluid stirrer in operation First stage compressor in operation Capillary cooling in operation Expansion valve cooling in operation Bath temperature Black temperature in tolerances set value 0 05 C Red temperature too high Blue temperature too low Heating in operation flashing ...

Page 18: ...setting and temperature corrections press Lock button If you forget to press Lock button after 2 minutes it locks automatic 4 2 4 Graph screen Press Graph button to see this screen Enter unlock code 0207 Press to Lock Press OK to confirm and unlock Enter full screen button Back to Main menu button Slider to move graph Slider to zoom graph in y direction Slider to move graph Slider to zoom graph in...

Page 19: ... parameters can seriously affect temperature control stability stability stabilization time This should only be done by skilled professional staff Set temperature Set each decimal place separately Set LCD brightness Next screen Increase value Decrease value Previous main screen Touch value to select Next screen Increase value Decrease value Touch value to select Previous screen ...

Page 20: ...ease value Touch value to select Previous screen Fluid mixing speed After changing this parameter wait for a few minutes for the system to stabilize Never change for more than a few at a time Set the heater power limit Set the frequency for the temp Averaging Next screen Increase value Decrease value Touch value to select Previous screen Temperature under which compressor starts cooling Temperatur...

Page 21: ...2 LT 21 4 2 9 Temperature corrections Default values Corr At 40 0 C 40 000 Corr At 0 0 C 0 000 Corr At 60 0 C 60 000 Corr At 130 0 C 130 000 Next screen Increase value Decrease value Touch value to select Previous screen ...

Page 22: ...le you will see this screen This will immediately stop temperature control The cooling system compressor will stop after some delay 4 2 11 Back panel Switch off the display Over temperature safety device shaft Over temp safety device reset spring Power supply cable Main switch RS 232 communication port ...

Page 23: ...e cleaned regularly to prevent a buildup of oil or dust For cleaning use some soft cleaning detergents and soft cloth Periodically check the fluid level in the bath to ensure that the level has not dropped A drop in the fluid level affects the stability and dramatically the uniformity of the bath Changes in fluid level are dependent upon several factors most often it is the volumetric expansion of...

Page 24: ...same time Large flow section is extremely important when replacing damaged fluid with high viscosity Needed tools to operate Pip grips such as Hexagon key 6mm such as 1 Allocate valve behind front removable deck 2 Remove isolation caps only for ULT models 3 Use hand or pipe grips to unscrew counterclockwise valve cap 4 Insert hexagon key 6mm and unscrew to open valve counterclockwise make sure to ...

Page 25: ...arest wall It must also be in a dry and well ventilated room Additional space or relocation of the apparatus may be necessary when carrying out any maintenance or servicing The user must provide electrical connection 230V 10 50Hz Environmental and operational conditions Temperature 15 C 25 C Relative humidity from 15 to 75 ...

Page 26: ...ssing a small spring next to the over temperature safety devices shaft Do to the large temperature range and possibilities of usage of different fluids there is a chance of fluid volume expansion at higher temperatures To avoid unlikable leakage the bath is equipped with overflow drain opening which leads to the lower part of the housing The operator must make sure that the catch pod is located di...

Page 27: ...duct with your other household waste Instead you should protect human health and the environment by handing over your waste equipment to a designated collection point for the recycling of waste electrical and electronic equipment For more information please contact your local waste disposal service ...

Page 28: ...eck stirring operating check fluid viscosity check cooling ribs on refrigerant system check temperature sensor Increased noise check stirring check fluid Not able to reach high temperature check the max power on heater limit setting Not able to reach low temperatures check and clean cooling ribs on cooling system check pressure in cooling system check stirring check fluid flow Probe error check pr...

Reviews: