37
0-10VDC Display Input Configuration
The 0-10VDC input is by far the most common display configuration. It has differential inputs
to the meter. It has a maximum resolution of 200 microvolts per count.
+/-5VDC Display Input Configuration
The +/-5VDC input option is used where a high resolution bipolar input is required. It has a
maximum resolution of 200 microvolts per count with the inputs configured single ended. This
option requires the input conditioning board (Kaman PN 853737-002) to be wired to the
display. This display is limited to a display reading of +/-9.99 (999x2 counts) since the first
digit location is taken up by the sign.
+/-10VDC Display Input Configuration
The +/-10VDC input option is used where a bipolar input with a large voltage swing is
required. It has a maximum resolution of 400 microvolts per count with the inputs configured
single ended. This option requires the input conditioning board (Kaman PN 853737-001) to be
wired to the display. This display is limited to a display reading of +/-9.99 (999x2 counts)
since the first digit location is taken up by the sign.