M A I N T E N A N C E
Under normal operation, the KAM OID should not require cleaning, unless pipeline usage is limited to a small num-
ber of products. Gasoline products or jet fuel in the pipeline will clean the OID without removal.
To remove any oil residues for visual inspection use a clean cloth with oil solvent or part washer.
Preferred solvents include, any petroleum solvent such as mineral spirits, isopropyl alcohol, gasoline, or diesel. Do
not use other chemicals.
If you have a question regarding cleaning solvents, please contact KAM CONTROLS directly at
+1 713 784-0000, or email: [email protected]
CLEANING AND INSPECTION
DIAGNOSTICS VIA RS232
Install RealTerm if not installed on your PC.
• RealTerm Software can be Downloaded at http://sourceforge.net/projects/realterm/files/Realterm/2.0.0.70/
• Click on Realterm_2.0.0.70._setup.exe to install.
Connect the RS232 cable to the OID as shown in the wiring diagram. FIG. 3-15 on page 15.
Connect the other end of the RS232 Cable to your computer serial port or to USB with the provided serial
converter.
Open RealTerm. A window will open up as shown in Fig. 5-1.
The program will open on the "Display" tab. Click on the up arrow of the "Cols" window until it reaches 120.
Do not try to type the number in the window as it will result in an error. If you do this, you must close the
program and start again.
1.
2.
3.
4.
5.
Connecting to the RS232 Serial Port using RealTerm
FIG. 5-1
Click until display
equals 120
Display tab