5
5. Installation
Do not deviate from the electric and temperature ratings printed on the data plate
located on the rear, lower right side of the unit.
CAUTION!
Incorrect operation of the equipment could result in potentially dangerous
conditions. To avoid hazards and minimize risk, follow all instructions and
operate the unit within the parameters stated on the data plate.
5.1 Positioning
Install the unit on a vibration free, level floor, with a
minimum
of fifteen centimeters
of space around the sides, rear, and top. Refer to Section 5.3 for instructions on
leveling the cabinet. Allow sufficient clearance for the door, or lid, to open at least 90
degrees.
Do not position the equipment in direct sunlight or near sources of heat such as
diffusers, radiators etc.
The ambient temperature range at the site must be b15
°
C to +32
o
C.
WARNING!
Install the unit on a solid, sturdy floor.
WARNING!
Never install the unit in a flammable or volatile environment. This
may result in fire or explosion.
Never install the unit where acid or corrosive gases are present.
Corrosion may result in wiring damage and current leakage/ electric
shock.
WARNING!
This unit is not designed for outdoor use. Current leakage or electric
shock may result if the unit is exposed to water.
Never install the unit in an excessively humid environment, or where
there is a high risk of being splashed by water. Deterioration of the