Kaise SK-7830 Instruction Manual Download Page 2

70-1201-7830-2 0807

Do  not  measure  any  current  that  might  exceed  maximum 
input value (200A DC / 600V line).
Read 

3.  SAFETY  PRECAUTIONS

carefully  to  avoid 

electric shock hazard and serious damage to the instrument.
Do not twist clamp head while measurement. Measurement 
should  be  incorrect  if  any  pressure  is  applied  to  the  clamp 
head.
Do  not  touch  any  part  of  power  line  or  the  circuit  to  be 
measured. 

Set 

FUNCTION

 Switch to "                 or                    ".

NOTE :

Do  not  touch  trigger  nor  open  clamp  head  until 
POWER-ON  INITIALIZE  is  completed  and  LCD 
indicates 0

±

1 digit.

Supporting Functions :  

Zero-Adjustment, Difference Measurement, Display Hold (Refer 
to 4-5 and  4-6).

When taking measurement in 4000mA range, to reduce 
geomagnetic  effect  to  the  measurement,  zero 
adjustment  by  POWER-ON  INITIALIZE  must  be  taken 
just  in  front  of  a  conductor  to  be  clamped,  fixing  the 
clamp head angle on measuring position.

NOTE  :

  LCD  indication  may  not  return  to  0mA  by 

geomagnetic effect if changing clamp head angle after 
measurement  taking  clamp  head  away  from  a 
measured conductor.
When measuring low current at 1000mA or lower, take 
zero  adjustment  by  POWER-ON  INITIALIZE  for  every 
measurement. 
LCD indication may not return to "0" by magnetic effect, 
if  high  current  is  instantaneously  applied  during  high 
current  measurement  at  100A  or  higher,  or  regular 
measurement in each range.
Under  4000mA  range  measurement, 

±

15mA  current 

fluctuation  from  measuring  value  is  not  displayed  on 
LCD by internal control.
Zero-point 

fluctuation 

occurs 

depending 

on 

temperature  alteration  of  measurement  environment 
that  could  cause  LCD  indication  not  to  return  to  "0"  or 
large measurement deviation.

NOTES : Following instructions are important to take  
                an accurate measurement.

4000mA  range  needs  a  few  seconds  until  LCD        
indications become stable.

Open  clamp  head  15mm  or  wider  and  close  it  softly.  Detach 
fingers from trigger after clamping a conductor.
Read the measurement value on LCD.

NOTE : 

After  measurement,  unclamp  from  the  conductor  and  set 

FUNCTION

 Switch  to 

"OFF"

.

 1.
 

 
 2.
 
 3.

 4. 

SK-7830 is warranted in its entirety against any defects of material 
or  workmanship  under  normal  use  and  service  within  a  period  of 
one  year  from  the  date  of  purchase  of  the  original  purchaser. 
Warranty  service  is  available  at 

KAISE  AUTHORIZED  SERVICE 

AGENCY 

through  your  local  dealer.  Their  obligation  under  this 

warranty  is  limited  to  repairing  or  replacing  SK-7830  returned 
intact or in warrantable defect with proof of purchase and transport 
charges  prepaid.   

KAISE  AUTHORIZED  DEALER

  and  the 

manufacturer, 

KAISE CORPORATION

,  shall not be liable for any 

consequential damages, loss or otherwise. The foregoing warranty 
is  exclusive  and  in  lieu  of  all  other  warranties  including  any 
warranty of merchantability, whether expressed or implied.
This  warranty  shall  not  apply  to  any  instrument  or  other  article  of 
equipment  which  shall  have  been  repaired  or  altered  outside  of 

KAISE  AUTHORIZED  SERVICE  AGENCY

,  nor  which  have  been 

subject to misuse, negligence, accident, incorrect repair by users, 
or  any  installation  or  use  not  in  accordance  with  instructions 
provided by the manufacturer.

Check  the  battery  connection,  polarity,  and  capacity  ("BAT" 
lights up or not).
Confirm that FUNCTION Switch is set to the correct position.
Confirm  that  measured  accuracy  is  adopted  in  the  operating 
environment.
Confirm  that  the  body  of  this  instrument  has  no  cracks  or  any 
other damages.

Periodical  check  and  calibration  is  necessary  to  make  safety 
measurements  and  to  maintain  the  specified  accuracy.  The 
recommended  check  and  calibration  term  is  once  a  year  and  after 
the repair service. This service is available at KAISE AUTHORIZED 
SERVICE AGENCY through your local dealer.

Repair  service  is  available  at  KAISE  AUTHORIZED  SERVICE 
AGENCY through your local dealer.  Pack the instrument securely 
with  your  name,  address,  telephone  number  and  problem  details, 
and ship prepaid to your local dealer.

KAISE AUTHORIZED DEALER

5-2. DC CURRENT MEASUREMENT (     mA /     A)

Dark Current : 

mA-level  low  current  that  is  used  after  turning  off  the  engine  by  such  as 
car  security  system  or  audio  settings  back-up.  Too  much  dark  current 
causes battery runs out, but its measurement was difficult. SK-7830 solved 
this problem and make it quick and easy.

Clamp-on a minus cable of car battery.
Read the measurement value on LCD.

NOTE : 

The values should be minus (

) if the current flows 

             opposite direction of "

" mark on clamp head.

If measurement value is higher than the specified value, check if 
any of car lamps remain lighting.

NOTE : 

POWER-ON  INITIALIZE 
does not work properly.

The  specified  dark  current  value  becomes  different  if  the 
optional  electric  components  such  as  car  navigation  or 
security  equipments  are  attached.  In  that  case,  refer  to 
their user's manuals for details. 

Measurement  value  should  be 
incorrect.

NOTE :

Especially for the cars that ECU 
or  other  electric  components 
keep  working  for  a  certain 
period  of  time  after  turning  off 
the  engine,  or  have  LCD  fuel 
meter.

Cannot measure dark current.
Turn the all lamps off.

Leave  the  car  engine  turned  off  for  about  15  minutes  after 
setting battery cable into the measurable condition.  All electric 
components  such  as  door  lamp,  room  lamp,  and  headlight 
must  be  turned  off.    Lock  the  all  doors  if  measuring  car  has 
remote-control door lock function. 

NOTE  :

  Leaving  time  is  different  depending  on  the  car 

functions,  especially  for  the  cars  that  ECU  or  other  electric 
components  keep  working  for  a  certain  period  of  time  after 
turning off the engine, or have LCD fuel meter.

NOTE :

 Cool down the engine before measurement, in case that 

measuring cable is close to the engine.

Put the clamp head close to a battery cable to be measured.  Set 

FUNCTION

  Switch  to 

"4000mA"

  without  touching  the  trigger.   

POWER-ON  INITIALIZE  works  to  adjust  LCD  display  into  "0" 
automatically.  (Refer to fig. 3)

WARNINGS

6-1. BATTERY REPLACEMENT

6-2. PERIODICAL CHECK AND CALIBRATION

6-3. REPAIR

WARNING

WARRANTY

 1.
 2.
 3.
 4.
 

 1.
 
 2.
 3.
 
 4.
 

Replace the batteries when "

BAT

" lights up on LCD.

Check the following items before asking repair service.

To avoid electrical shock, detach instrument from circuit when 
to replace battery. Set 

FUNCTION

 Switch to "OFF".

Turn the power off.
Unscrew the Baterry Cover and remove exhausted batteries.
Insert 2 of new 1.5V R6P (AA) batteries in correct polarity.
Fix battery Cover and tighten the screw.

422  Hayashinogo, Ueda City, Nagano Pref., 386-0156 Japan
TEL : +81-268-35-1600 (REP.)  /  FAX : +81-268-35-1603
E-mail : [email protected]
http://www.kaise.com

fig.9 

6

MAINTENANCE

6

7

8

5

4000mA

40A / 200A

● 

MEASUREMENT EXAMPLE 1.  

  

Automobile Dark Current Measurement

● 

MEASUREMENT EXAMPLE 2.  

● 

Example of Incorrect Measurement

Measurement of Car Alternator's Charging Current

Turn the power on with clamp head opened.

Start measurement with any of car lamps remain lighting.

Start measurement as soon as engine stopped.

R6P(AA)

Screw

Battery Cover

Two 1.5V R6P Batteries.

1.5V

1.5V

KAISE  JAPAN

4201-7830-1  0708

WARNING

MAX.     200A
BATTERIES : 1.5V     R6P

×

2  26mA MAX.

     OVERVOLTAGE  CAT.

 300V   CAT.   600V

SERIAL NO. 

TO  AVOID  ELECTRICAL  SHOCK, 
DETACH INSTRUMENT FROM CIRCUIT 
WHEN TO REPLACE BATTERY.

Trigger


Start the engine, and read the measurement value on LCD.

NOTE : 

The value should be minus (

) if the current flows 

             opposite direction of "

" mark on clamp head.

Alternator  has  no  problem  if  20A  to  40A  is  displayed  first,  and 
then it slowly becomes lower.
Alternator  should  be  defective  if  no  value  is  indicated,  or  high 
current value remains indicated.

Stop the car engine.

Set 

FUNCTION

 Switch to "        

   

 " .

Be sure not to touch trigger. (Refer to fig. 8)
Clamp-on B-terminal cable from car alternator.

fig.3

fig.4

fig.8

fig.5

fig.6

fig.7

4000mA

Product  specifications  and  appearance  are  subject  to  change  without 
notice due to continual improvements.

Car Alternator : 

Automobile  engine  generator  that  outputs  DC  electricitiy.  Measuring  its 
charging current is effective to find the trouble that might cause battery-runs-
out or battery-damages.

40A / 200A

1.

2.

3.

4.

5.