Page 16
Instruction manual - G95
Kivex Biotec Ltd
TUNE
The TUNE option is for
calibration
of the
displayed temperature.
If there is an offset between the value on the display and any measurements made with a high precision
external temperature sensor, this can be corrected.
The new temperature reading will be kept as the displayed value, and temperature controlling conduct-
ed on this basis.
When the power (incl. battery) is disconnected default values will be restored.
Follow these steps to navigate in the TUNE option.
Action
Key
Display
Press and hold UP+DOWN key for 3
seconds to enter the the menu.
The option "UNIT" will appear in the
display.
Press DOWN key until the option "TUNE"
appears in the display.
Press and hold setpoint key to change
values.
While holding the setpoint key in, press
either the arrow UP or arrow DOWN key to
select the value of temperature.
When the required temperature has been
selected let go of the setpoint key.
Press and hold UP+DOWN key for 3
seconds to exit the the menu.
To see the calibrate value
Press and hold setpoint key.
While holding the setpoint key in, press the
HEAT key
The calibrate value
is shown
CAUTION
Any change in the factory set calibration should always be based on very certain
temperature measurements. K-SYSTEMS only recommend the use of a high quality
PT-100 sensor that is calibrated with the precision instrument used for measuring
according to the manufacturers specifications. Also ensure an optimal contact between
the sensor and the place measured. Use only a sensor type that is correct for the
purpose and of correct size. If in doubt contact your local distributor or K-SYSTEMS.