90.0~109.9°F (32.0~43.9°C)
0.1°F (0.1°C)
95.9~107.6°F
:
±0.2°F, Other
:
±0.3°F
(35.5~42.0°C
:
±0.1°C, Other
:
±0.2°C)
Last memory
LR41 Battery x 1pcs
More than 5000 measurements
123.7mmx18.1mmx10.4mm
Approx. 9.8g includes battery
60.8~104.0°F(16~40°C), R.H. of 15~95% non-condensing
-4~122°F(-20~50°C), R.H. of 15~95% non-condensing
ISO 13485
ASTM E1112
This digital thermometer is intended to measure human body temperature for children and adults; it has digital
display and can be used in household environment.
The product is contraindicated:
1. Place the thermometer probe in scarred tissue or tissue compromised by any disorders of measuring site.
2. Place the thermometer probe on skin exposed to direct sunlight, fireplace heat, cold compress therapies, air
conditioner flow, etc.
1. Easy to read digital LCD (liquid crystal display).
2. Memory display for last temperature taken.
3. Compact, accurate and durable LSI (large scale integration) device.
4. Very sensitive to obtain temperature.
5. If the thermometer is inadvertently left on after temperature stabilization, it will automatically shut off in 3 minutes.
6. Small light weight unit.
7. When low battery indication shows, replace the old battery with a new battery.
8. The unit is water resistant.
Measuring Range
Resolution
Accuracy
Memory
Battery(include)
Battery Life
Dimensions
Weight
Operating Condition
Storage Condition
Certification
Compliance
1. Do not boil the probe. Instead, clean the unit by wiping it with a dry cloth and disinfect the probe with isopropyl
(rubbing) alcohol. See cleaning instructions for details.
2. Do not drop the thermometer or expose it to heavy shock.
3. Do not bite or swallow the probe and the battery cover.
4. Do not let thermometer come into contact with any thinner or chemical solvents.
5. After using, clean the thermometer and place it into the carrier case.
6. Do not store the unit under direct sunlight, at a high temperature, in high humidity or dust. Performance may be
degraded.
7. Stop using the thermometer if it operates erratically or if the display malfunctions.
8. Keep out of reach of unsupervised children. Do not let children play with the unit. TEMPERATURE SHOULD BE
TAKEN UNDER ADULT SUPERVISION.
9. Components may be harmful if swallowed. Contact a physician immediately should this occur.
10. Clean the thermometer probe before storing.
11. Do not attempt to disassemble the unit except to replace the battery.
12. Do not use if the display malfunctions or it operates irregularly.
13. Do not put this unit in dishwasher.
14. If the thermometer has been stored at below-freezing temperatures, allow the thermometer to warm naturally to
room temperature before using.
15. Performance of the thermometer may be degraded if: Operated or stored outside stated temperature and humidity
ranges or if the patient's temperature is below the ambient (room) temperature.
16. Before using, it is recommended to no drinking, smoking or excising for at least twenty minutes prior to do a
measurement and no talk, walk during measuring, it may cause inaccurate reading.
17. When the battery is almost exhausted, the symbol will appear in the display window, meaning it’s time to
replace the battery.
18. ORAL TEMPERATURES READLINGS ARE OFTEN 1.0°F LOWER THAN ACTUAL BODY TEMPERATURE
READINGS.
19. The readings is for reference, if feel sick or uncomfortable, please contact the physician.
1. When the symbol “ ” appears, the battery needs to replaced.
2. Pull the battery cover off in the direction shown.
3. Use a pointed object such as a toothpick or insulated probe to pry loose the old battery. Avoid of using any sharp
metal object in this operation.
4. Place a new 1.5V button size battery, LR41 or equivalent in the chamber with the po) side up and the
negative(-) side down.
5. Replace and fasten the battery cover securely.
1. Discard old battery carefully, out of reach of young children. Swallowing the battery may be fatal. If the battery is
swallowed, contact a hospital immediately to have it removed. Do not dispose of the battery in a fire. It may explode.
2. For battery disposition, please consider the national regulation.
3. If no use for 1 month, please take out the battery to avoid the battery deterioration.
DEVICE DESCRIPTION AND INTENDED USE
CONTRAINDICATIONS
Low Battery Indication
Memory Indication
Measure Value
Measure Unit
DIGITAL CLINICAL THERMOMETER
PARTS DESCRIPTION
SPECIFICATIONS
PRECAUTION
BATTERY REPLACEMENT
Warning
FEATURES
KD-112
1. Disinfect the probe with isopropyl (rubbing) alcohol before using.
2. Press and release the ON/OFF button, the display read confirming that the unit is functioning properly.
3. Then the display will show L°F with °F flashing. If the room temperature is greater than 90.0°F, it will display rather
than L°F.
4. Place the probe in the appropriate position.
5. Once the degree sign (°F) stopped flashing, the measurement is finished and the temperature is displayed. An
alarm signal will also sound for approximately 8 seconds after °F stop flashing.
6. The unit will automatically turn off in 3 minutes (approx.). However, to prolong battery life, it is best to press the
ON/OFF button to turn off once the temperature has been indicated.
Special Features:
Last-Temperature Memory Recall: When the unit is off, press and hold the ON/OFF button around 3 seconds, the last
memory will be shown. Release the ON/OFF button, the unit will be ready for measuring.
For Models with Fever alarm: If the temperature registers higher than 99.5
℉
, the fever alarm (a quick beeping sound
which occurs continuously for about 8 seconds) will sound when the peak temperature has registered. Also, the result
will keep flashing until the thermometer is turned off.
For Switchable Centigrade and Fahrenheit Models: After turning on the thermometer, the Centigrade or Fahrenheit
mode can be switched within first 5 seconds by pressing and holding the ON/OFF button for 2 seconds.
HOW TO USE
Probe
Display Window
ON/OFF Button
Battery Cover