19
limitations on the spacing between trigger events in each mode. Trigger events arriving too
close to the previous instance will be ignored by the digitizer.
• In Post-Trigger mode, the minimum spacing between trigger events is N+1
• In Delayed-Trigger mode, the minimum spacing between trigger events is (N+D)+1,
where D is the number of the delayed setting
Figure 3-10: Re-Trigger Mode Acquisition