Capacitance Meter Manual                                                           DN060-02v04 

JYE Tech                                  - 2 -                               www.jyetech.com

 

 

Note: When we say voltages we mean voltage with refer to GND if not stated otherwise. 

2.

 

The display is stuck at “C0” when zeroing button is pressed 

Ø

 

Check R11, R13, R15, R16, and related connections.   

Ø

 

Test poles short (pin 1, 2, and 3 short to GND) can cause this problem too. 

3.

 

Broken Display 

Ø

 

Check all connections related to the LED display. Look for opens as well as shorts. 

4.

 

Bad serial data output 

Ø

 

Make sure level converter is in place if connected to standard RS232 devices. 

Ø

 

Make sure correct data format is selected. 

 

 

Revision History 

Version 

Date 

Descriptions 

01 

2008.09.17 

First created 

02 

2008.11.28 

Modified for 109-06000-00B PCB. Removed reworking instructions since errors in 
previous PCB had been corrected.   

03 

2009.12.01 

Added more detailed explanation for serial date format. 

04 

2010.02.25 

Re-wrote the section  “How To Assembly”.   Revised the section  “How to Use”. 
Added the section “Troubleshooting” 

 

 

 

 