Capacitance Meter Manual DN060-02v04
JYE Tech - 2 - www.jyetech.com
Note: When we say voltages we mean voltage with refer to GND if not stated otherwise.
2.
The display is stuck at “C0” when zeroing button is pressed
Ø
Check R11, R13, R15, R16, and related connections.
Ø
Test poles short (pin 1, 2, and 3 short to GND) can cause this problem too.
3.
Broken Display
Ø
Check all connections related to the LED display. Look for opens as well as shorts.
4.
Bad serial data output
Ø
Make sure level converter is in place if connected to standard RS232 devices.
Ø
Make sure correct data format is selected.