Troubleshooting
If abnormalities occur during FreedomView
®
SD LED Fiberscope Inspection Kit operation, discontinue use and
consult Optim, Incorporated.
DO NOT ATTEMPT SELF-REPAIR!
IMAGE IS NOT CLEAR
POSSIBLE
CAUSE
SOLUTION
Eyepiece diopter not adjusted
Adjust diopter ring until fiber pattern
to user’s vision:
is in clear, sharp focus.
Target is out of focus:
Reposition objective end to focal
distance
of
6mm-100mm.
Contaminant on lens:
Clean lens (Refer to Section 3G).
Fluid leakage into the internal shaft:
Return to OPTIM for service.
Loose Lenses:
Return to OPTIM for service.
INSUFFICIENT ARTICULATION
POSSIBLE
CAUSE
SOLUTION
Stretched
control
wires:
Return to OPTIM for service.
Loose
lever:
Return
to
OPTIM
for
service.
Shaft
Damage:
Return
to
OPTIM
for
service.
Excessive number of bends in shaft:
Access path may be too complicated or
intricate.
NO ILLUMINATION
POSSIBLE
CAUSE
SOLUTION
Battery
is
discharged:
Refer
to
Section 3, Lithium Ion Battery, Battery Charger,
And
AC
Power
Supplies,
for
instructions
USE THE POWER CONVERTERS SUPPLIED WITH THE FreedomView® SD LED Fiberscope
KIT ONLY.