EN
26
Guidance and manufacturer's declaration – Electromagnetic
immunity –for all equipment and systems
Guidance and manufacturer´s declaration – Electromagnetic immunity
The Infrared Thermometer JPD-FR202 is intended for use in the
electromagnetic environment specified below. The customer or the user of the
Infrared Thermometer JPD-FR202 should assure that it is used in such an
environment.
Immunity test
IEC
60601
test level
Compliance
level
Electromagnetic
environment- guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6
kV
contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic
material,
the
relative humidity should be
at least 30 %.
Power
frequency
(50/60 Hz)
magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Guidance and manufacturer´s declaration – Electromagnetic
immunity –for equipment and systems that are not life-supporting
Guidance and manufacturer´s declaration – Electromagnetic immunity
The Infrared Thermometer JPD-FR202 is intended for use in the
electromagnetic environment specified below. The customer or the user of the
Infrared Thermometer JPD-FR202 should assure that it is used in such an
environment.
Immunity test
IEC
60601
test level
Compliance
level
Electromagnetic
environment -guidance
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz
to 2.5
3 V/m
Portable and mobile RF
communications
equipment should be used no