Signal processing: special digital signal processing unit
Measuring range: 60bpm to 210bpm
Alarm Range:
Upper limit: the adjustable range is 150bpm to 180bpm, the default is 160bpm
Lower limit: the adjustable range is 90bpm to 120bpm, the default is 110bpm
Maximum audio output power: 2W
UC pressure
Measuring range: 0-100mmHg
Display
The LCD screen shows the monitoring state of FHR curve, UC pressure curve, FM symbols, volume and so on,
which can be frozen and replayed.
Power supply: AC 100-240V 50/60Hz, internal battery: lithium battery3.7V 3000mAh
Input Power: 20VA
Work mode: continuous working for more than 10 hours after full power supply
Size: 89.2mm (length)*202mm(width)*75.9mm (height)
Net weight: 882.5g (power adapter and accessories are not included)
Environment
Working environment: temperature: +5
℃
to +40
℃
humidity:
≤
80% atmospheric pressure: 86KPa to 106KPa
Transportation and storage environment: temperature: -20
℃
to +55
℃
humidity:
≤
95% atmospheric pressure:
50KPa to 106KPa
Probe sound output: According to the regulations of international standard IEC 61157, the peak negative acoustic
pressure shall not exceed 1 MPa, and the output beam sound intensity shall not exceed 20mW/cm
2
. The