Fetal Doppler

11

Share your Recording on Social Media

Share the recording you made with your family and friends on your own
Social Media like Facebook, Twitter.

Principles of how the Fetal Doppler works

This Fetal Doppler consists of a receiver unit that can detect the slightest of

sounds inside the womb and sends signal output to a loudspeaker or a

smartphone.

A sound wave is transmitted from the Fetal Doppler to the uterus in the mommy's

belly and an echo is then received back when the sound wave bounces off of the

baby's heart.

The difference between these two signals (sound waves) are measured to

determine the baby's heartbeat.

Cleaning, Disinfecting, Care and Disposal

We recommend cleaning the gel from the transducer probe with a soft cloth. You
can disinfect the probe by wiping it with 70% ethanol or alcohol wipes and then
drying with a soft dry cloth. Do not use abrasives, solvents, bleach, soap and water,
steam or high temperature cleaning processes to sterilize the unit as any of these
can damage it. Do not immerse in water or liquids to clean the Fetal Doppler as this
will short out the unit.

The battery is recyclable. Please recycle the battery after use by following your
local recycling guidelines.

Summary of Contents for AngelSounds JPD-100S9

Page 1: ...JPD 100S9 Fetal Doppler USER MANUAL Copyright Jumper Medical All rights reserved Manual Version Version 1 1 Issuing Date 2020 11...

Page 2: ......

Page 3: ...s Email info jumper medical com Website http www jumper medical com MedPath GmbH Mies van der Rohe Strasse 8 80807 Munich Germany Please read the User Manual carefully to make sure safe and proper use...

Page 4: ...arance 3 Controls and Indicators 4 Location and Detection 8 Recording your Baby s Heartbeat 10 Share your Recording on Social Media 11 Principles of how the Fetal Doppler works 11 Cleaning Disinfectin...

Page 5: ...then troubleshoot before trying to find the fetal heartbeat 4 When the Fetal Doppler is working correctly you can now lay back and apply coupling agent to the surface of the ultrasonic probe 5 Now pu...

Page 6: ...y s heartbeat The normal range of fetal heart rate 110bpm 160bpm CAUTION It cannot replace the professional fetal monitor when the fetal heart rate is abnormal or can not find the fetal heart can not...

Page 7: ...ce used for detecting fetal heartbeats Note for home use This device cannot replace a professional fetal monitor If the fetal heart rate is abnormal or cannot be located using this monitor immediately...

Page 8: ...d to change the volume Press the Volume up button the sound volume will increase Contrary press the Volume down button the sound volume will decrease Recording jack Used to connect a smartphone via a...

Page 9: ...and matched 3 Close the battery cover as the bellow picture shows Additional Battery Notes Be sure to remove the battery s protective clear wrapper before installing it in the Fetal Doppler If the Fe...

Page 10: ...tore or Google Market by entering the keyword JUMPER Health Notes This APP supports iOS 7 0 and later versions and Android 4 3 and later versions In addition hardware of the smartphone needs to suppor...

Page 11: ...on the belly Baby Oil Coconut Oil Virgin Olive Oil or even Vaseline work well which is also readily available in most homes Actually anything that lubricates the movement across your belly will work...

Page 12: ...omen to listen to the baby s heartbeat which should be much faster than mommy s heartbeat 6 Move the Fetal Doppler very very slowly to find the optimum placement based on the baby s current position 7...

Page 13: ...a break and trying again a few days later However if you are concerned about any issues relating to your pregnancy you should contact your healthcare provider for assistance As the baby grows the hea...

Page 14: ...Fetal Doppler is a listening system to hear the sounds your baby makes and is NOT a medical device and can not be used for diagnostic or medical purposes and is not to be used as a substitute for regu...

Page 15: ...ces off of the baby s heart The difference between these two signals sound waves are measured to determine the baby s heartbeat Cleaning Disinfecting Care and Disposal We recommend cleaning the gel fr...

Page 16: ...eak battery can cause noise Interference can be caused by nearby electronic devices e g high frequency electronics cell phones cordless phones etc Make sure there is plenty of coupling gel between the...

Page 17: ...sier to hear the heartbeat when you find the right location The baby can sometimes be located behind the placenta making it difficult to find try again later The battery cover doesn t want to stay on...

Page 18: ...equipment P_ 1MPa Iob 20mW cm2 Ispta 100mW cm2 Audio output power 0 2 W Audio out jack 3 5mm Recommended battery type 9V DC alkaline battery Stand by Time 4hours Suitable gestation 16 gestational wee...

Page 19: ...tion against ingress of water and particulate matter It indicates that the equipment should be sent to the special agencies according to local regulation for separate collection after its useful life...

Page 20: ...suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domesticpurposes A...

Page 21: ...in the electromagnetic environment specified below The customer or the user of the Fetal Doppler should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance leve...

Page 22: ...rption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM...

Page 23: ...er of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7...

Page 24: ...Shenzhen Jumper Medical Equipment Co Ltd Address D Building No 71 Xintian Road Fuyong Street Baoan Shenzhen Guangdong China Tel 86 755 26696279 Fax 86 755 26852025 Website http www jumper medical com...

Reviews: