6 Annex
82
6.1.9
Environmental influences
6.1.10
Case
Conductor cross section on termi-
nals 6 to 23
Wire or stranded wire
without ferrule
Min. 0.14 mm
2
, max. 1.5 mm
2
(with terminal cover: max. 0.5 mm
2
)
Stranded wire with ferrule
Without plastic collar: min. 0.25 mm
2
, max. 1.5 mm
2
(with terminal cover: max.
0.5 mm
2
)
With plastic collar: min. 0.25 mm
2
, max. 0.5 mm
2
Stripping length
9 mm
Voltage supply influence
< 0.1 % of the measuring range
Ambient temperature range
Storage
-20 to +60 °C
Operation
0 to 50 °C; in conjunction with extra code 970: 0 to 40 °C
Site altitude
Max. 2000 m above sea level
Climatic environmental influences According to DIN EN 60721-3 with extended temperature range
Resistance to climatic
conditions
≤
85 % rel. humidity without condensation
Storage
According to class 1K2
Operation
According to class 3K3
Mechanical environmental
influences
According to DIN EN 60721-3
Storage
According to class 1M2
Transport
According to class 2M2
Operation
According to class 3M3
Electromagnetic compatibility
(EMC)
According to DIN EN 61326-1
Interference emission
Class A – only for industrial use –
Interference immunity
Industrial requirements
Case type
Flush-mounted housing according to DIN IEC 61554 made of zinc-plated steel
sheet (indoor use)
Case front
Made of die-cast zinc with decor foil
Front frame dimensions
144 mm x 144 mm (front frame depth approx. 8 mm including seal)
Mounting depth
120.9 mm (incl. spring-cage terminals)
Panel cut-out
138
+1.0
mm × 138
+1.0
mm
Panel thickness
2 to 8 mm
Case fastening
In panel using the four supplied mounting elements
Operating position
Any, with due consideration for the viewing angle of the screen,
horizontal ±50°, vertical ±30°
Protection type
According to DIN EN 60529, front IP65, back IP20;
in conjunction with extra code 970: IP20 with open carrying case, IP20D with
closed carrying case
Weight
Max. 1.75 kg (without terminal cover)