7 Commissioning
7.1
Getting started
✱
Mount the device, see chapter 4 "Assembly", page 12.
✱
Install the device, see chapter 5 "Installation", page 13 ff.
✱
Call up Administrator level (ADMINISTR. LEVEL).
✱
Enter password 0300 (factory setting).
✱
Call up PARAMETER LEVEL/DISPLAY/OPERAT. TIMEOUT.
✱
Set OPERAT. TIMEOUT to 0 minutes (no timeout).
✱
Leave the Display level with "EXIT"
✱
Leave the Parameter level with "EXIT"
✱
Select BASIC SETTING and work through all the menu items, see chapter
6.7.3 "Basic settings", page 35.
✱
Answer "YES" to the "Reinitialize device" query
✱
Configure the required additional parameters.
✱
Calibrate the device to the sensor and sample medium,
see chapter 8 "Calibrating a pH measurement chain", page 51 or
see chapter 9 "Calibrating a redox measurement chain", page 60 or
see chapter 10 "Calibrating an ammonia sensor", page 67 or
see chapter 11 "Calibrating a sensor with a standard signal", page 70.