39
6 Operation
Control is not longer via the device. The output level of the controllers is 0 %.
Controller 1 is activated by the
key. In this case the output level of
controller 1 is 100 %.
Controller 2 is activated by the
key. In this case the output level of
controller 2 is 100 %.
Deactivation
✱
Press the
key.
Control is once again through the outputs of the device.
The word MANUAL appears in the status line of the display.
6.8.2
Simulation of binary outputs
Activate simulation
✱
Set ADMINISTR. LEVEL/PARAMETER LEVEL/BINARY OUTPUTS/
BINARY OUTPUT1 ( ... 8) "Manual mode no simulation, Inactive or Active".
No simulation
= No Manual mode, control is via device.
Inactive
= Relay K1 or K2 is de-energized;the word MANUAL
appears in the status line of the display
Active
= Relay K1 or K2 is energized;the word MANUAL appears
in the status line of the display
Deactivate manual mode
No simulation = No Manual mode, control is via device.
When the device is in display mode, the word MANUAL disappears from the
status line of the display.
If the
keys (alone) are pressed for longer than 3 seconds, the device
switches to language selection!
If the
and
keys are pressed for longer than 3 seconds, the device goes
into HOLD mode.
Then the outputs of the device respond according to the default settings.
To exit HOLD mode, press the
and
keys for longer than 3 seconds.
EXIT
EXIT
EXIT
EXIT
In the factory setting of the device the MANUAL mode parameter is set to "No
simulation" and can only be activated by the administrator!
This parameter must first be released for other users, see "Release level",
page 35.
If a higher order switching function has been assigned to an output, Simulation
mode is not possible for that output.