4 Electrical connection

4

Electrical connection

4.1

Installation notes

4.2

Connection diagram

Type 71 with fixed cable, 6 × 0.14 mm, shielded

Type 70 with M12 connector, 8-pole, A-coded

Route input, output, and supply lines separately and not parallel to one another

To connect the conductivity sensor, use the CR 4P-cable (see "Accessories") if possible; do not run 
the connecting cable close to current-carrying components or cables

Implement the cables continuously (not via terminal blocks or similar)

Do not connect any additional loads to the supply terminals of the sensor

Follow the instructions regarding the electrical connection in the operating manual for the transmitter 
used

Make sure that you always use safety devices that are independent of the device, and ensure that 
only qualified personnel configure the settings. In addition to incorrect installation, incorrectly set val-
ues on the transmitter can impair the performance of the downstream process or cause damage

Figure

Wire color

Function

Green

Pt1000

Yellow

Pt1000

Gray

Inner electrode 2

Brown

Outer electrode 2

White

Outer electrode 1

Pink

Inner electrode 1

Figure

PIN

Function

1

Pt100/Pt1000

2

Pt100/Pt1000

3

Inner electrode 2

4

Outer electrode 2

5

Pt100/Pt1000 three-wire circuit 

6

Outer electrode 1

7

Inner electrode 1

8

Not used

Summary of Contents for 202931

Page 1: ...20293100T90Z001K000 V2 00 EN 00743106 2020 08 13 JUMO BlackLine CR 4P Conductive 4 electrode conductivity sensor Type 202931 Operating Manual...

Page 2: ......

Page 3: ...g the device version 8 2 1 Nameplate 8 2 2 Order details 8 2 3 Accessories 9 3 Mounting 10 3 1 General information 10 3 2 Installation suggestions 10 4 Electrical connection 11 4 1 Installation notes...

Page 4: ...4 1 Introduction 1 Introduction...

Page 5: ...cautionary measures are not carried out CAUTION This symbol in connection with the signal word indicates that material damage or data loss will occur if the respective precautionary measures are not t...

Page 6: ...and redox electrodes in corre sponding combination fittings The sensor is also particularly suited for connection to hand held and laboratory measuring devices 1 4 Areas of application Thanks to its b...

Page 7: ...nally excitation and measurement are kept separate This has some advantages compared to 2 electrode measuring sensors as the effects of polarization recede into the background To a large extent incomi...

Page 8: ...above the sensor was produced in 2020 in week 30 3 Cell constant K This is where the nominal cell constant of the sensor is specified 2 2 Order details 1 Basic type 202931 JUMO BlackLine CR 4P Conduct...

Page 9: ...4P cable 10 m assembled with M12 connector 00528700 CR 4P cable 5 m assembled with M12 connector version for connection to JUMO AQUIS touch extended connecting cable for Pt1000 00502027 CR 4P cable 10...

Page 10: ...g material Insertion length The insertion length of the conductivity sensor must be coordinated with the installation situation 3 2 Installation suggestions The sensor can be installed in the followin...

Page 11: ...minals of the sensor Follow the instructions regarding the electrical connection in the operating manual for the transmitter used Make sure that you always use safety devices that are independent of t...

Page 12: ...rollers M12 cable socket Metal 8 pole Cable diameter 7 5 mm Cable material PUR blue Ambient temperature 30 to 80 C Cable length 5 m or 10 m YE 7 5 GN BN BL GR PK RD Type Features Data sheet JUMO dTRAN...

Page 13: ...Settings on the transmitter controller NOTE For the calibration it is essential that you observe the operating manual of the transmitter controller used e g JUMO AQUIS 500 CR JUMO dTRANS CR 02 JUMO AQ...

Page 14: ...odes in an ultrasonic bath is also good for removing various deposits When cleaning proceed as follows Wipe off minor dirt with a soft preferably lint free cloth Wash off stubborn dirt and greasy or o...

Page 15: ...sor dirty Clean conductivity sensor Transmitter does not show con ductivity e g display 0 Line breakage or incorrect elec trical connection Check electrical connection Sensor in air not immersed in me...

Page 16: ...pressure 6 bar at 80 C Protection typee IP65 and IP67 Electrical connection Either securely connected cable 5 m long with tin plated wire ends or M12 connector a The manufacturing related deviation o...

Page 17: ...QHFWLRQ 3UR HVVDQVFKOXVV 1XWV 0XWWHU 6FUHZ 6FKUDXEH d 7KLV WDEOH LV SUHSDUHG LQ DFFRUGDQFH ZLWK WKH SURYLVLRQV 6 7 7 QGLFDWH WKH KD DUGRXV VXEVWDQFHV LQ DOO KRPRJHQHRXV PDWHULDOV IRU WKH SDUW LV EHORZ...

Page 18: ......

Page 19: ......

Page 20: ...ers Road East Syracuse NY 13057 USA Delivery address Mackenrodtstra e 14 36039 Fulda Germany Phone Fax Email Internet 44 1279 63 55 33 44 1279 62 50 29 sales jumo co uk www jumo co uk Phone Fax Email...

Reviews: