![Julabo HL-4 Operating Manual Download Page 26](http://html1.mh-extra.com/html/julabo/hl-4/hl-4_operating-manual_2030693026.webp)
Safety installations, warning functions
26
8.
Safety installations, warning functions
Check the safety installations at least twice a year! Refer to page 13.
Settings for the excess temperature protection > SafeTemp< and for the
warning functions > OverTemp< and > SubTemp< are made in a menu
which is called up by pressing the key
.
Menu item > Type< allows choosing between a warning and an
alarm cut-off for the menu items > OverTemp< and > SubTemp<.
Warning:
The excess temperature protection
m u s t
be set at least 25 °C below the fire point of
the bath fluid used!
In case of wrong setting there is a fire hazard!
We disclaim all liability for damage caused by wrong settings!
SafeTemp:
Setting range: 20 °C ... 320 °C
Rough setting can be effected by using the
temperature scale.
Settings:
Press the key
and by pressing the key
or
select
the menu item
and confirm the by pressing the key
.
For setting proceed as follows.
SafeTemp:
Set the new cut-out value within 30 seconds with a screw driver.
Exact display and setting on LCD display
Recommendation:
Set over temperature-protective setting >SafeTemp< 5 °C to 10 °C
above the set point of the working temperature.
Type:
Switch-over from warning to shutdown function
Change the parameter by pressing the key
and confirm by
pressing the key
.
or
pressing the key
if the parameter is to be retained.
Possible parameters:
warning or alarm
OverTemp :
SubTemp:
Set value on numeric keypad and confirm by pressing
Example: 83.00 °C
Press the keys
and
and confirm by pressing