18
4.3. Safety
instructions
Follow the safety recommendations to prevent damage to persons or property.
Further, the valid safety instructions for working places must be followed.
Only connect the unit to a power socket with earthing contact (PE – protective
earth)!
The power supply plug serves as a safe disconnecting device from the line and
must always be easily accessible.
Place the instrument on an even surface on a pad made of non-inflammable
material.
Do not stay in the area below the unit.
Make sure you read and understand all instructions and safety precautions listed
in this manual before installing or operating your unit.
Never operate the unit without bath fluid in the bath.
Do not drain the bath fluid while it is hot or cold!
Check the temperature of the bath fluid prior to draining (by switching the unit on
for a short moment for example).
Use suitable connecting tubing.
Make sure that the tubing is securely attached.
Avoid sharp bends in the tubing, and maintain a sufficient distance from
surrounding walls.
Regularly check the tubing for material defects (e.g., for cracks).
Never operate damaged or leaking equipment.
Always turn off the unit and disconnect the mains cable from the power source
Summary of Contents for FL300