Start:
•
Press the start/stop key.
- The MULTI-DISPLAY (LED) indicates the actual bath
temperature. (example: 21.0 °C)
- An illuminated indicator light in the "T1" or "T2" key indicates
the activated working temperature.
Stop:
•
Press the start/stop key.
The MULTI-DISPLAY (LED) indicates the message "OFF".
The unit also enters the safe operating state "OFF" or "r OFF after a mains power
interruptance. In case of manual operation, the start/stop key has to be pressed to
restart operation. The temperature values which were entered via the keypad of the
circulator remain in memory.
If the circulator is operated in remote control mode, the order to start and all values
which have to be adjusted, must be resent via interface by the personal computer.
NOTE:
The circulator has been configured and supplied by JULABO according to N.A.M.U.R.
recommendations. This means for the start mode, that the unit must enter a safe operating state after a
power failure (non-automatic start mode). This safe operating state is indicated by „OFF“ or „rOFF“,
resp. on the MULTI-DISPLAY (LED). A complete shutdown of the main functional elements such as
heater and circulation pump is effected simultaneously.
Should such a safety standard not be required, the AUTOSTART function (automatic start mode) may
be activated, thus allowing the start of the circulator directly by pressing the mains power switch or using
a timer.
Automatic / non-automatic start mode
Activating/deactivating AUTOSTART
Press enter
and the start/stop key
at the same
time,
and turn on the circulator with the mains power switch.
For a short while the MULTI-DISPLAY indicates the effective start
mode:
AUTOSTART on.
AUTOSTART off.