MC
49
F
The
out
commands are valid only in remote control mode.
Examples:
Command to set the working temperature >t 1< to 55,5 °C:
out_sp_00
⇔
55.5
↵
Command to ask for the working temperature >t 1<:
in_sp_00
↵
Response from the circulator:
55.5
↵
11.3. List of commands
out commands:
Setting parameters or temperature values.
Command
Parameter
Response of circulator
version
None
Number of software version (V X.xx)
status
none
Status message, error message (see page 51 )
out_mode_01
0
Use working temperature >t 1<
out_mode_01
1
Use working temperature >t 2<
out_mode_01
2
Use working temperature >t 3<
out_mode_05
0
Stop the unit = R –OFF-.
out_mode_05
1
Start the unit.
out_sp_00
xxx.xx
Set working temperature. „t 1“
out_sp_01
xxx.xx
Set working temperature. „t 2“
out_sp_02
xxx.xx
Set working temperature. „t 3“
out_sp_03
xxx.xx
Set high temperature warning limit
„t High“
out_sp_04
xxx.xx
Set low temperature warning limit
„t Low“
out_sp_07
x
Set the pump pressure stage. (1 ... 4)
out_par_06
xxx
Xp control parameter of the internal controller.
out_par_07
xxx
Tn control parameter of the internal controller.
out_par_08
xxx
Tv control parameter of the internal controller.