
Menu functions
42
>
tYP
< - ATC TYPE
•
The parameter flashes, switch by pressing
and
L
A
>1-point<, >2-point<
or
>3-point<
calibration
can be carried out.
The value >
tt1
< is only indicated
L
In addition the measured temperature value
>
Ct 1
< is saved during the next step.
•
Integer digits flash, set by pressing
+
•
Decimal digits flash, set by pressing
+
L
If only a 1-point calibration is carried out, the following menu items
are not indicated anymore
The value >
tt2
< is only indicated
L
In addition the measured temperature value
>
Ct 2
< is saved during the next step.
•
Integer digits flash, set by pressing
+
•
Decimal digits flash, set by pressing
+
L
If only a 2-point calibration is carried out, the following menu items
are not indicated anymore
The value >
tt3
< is only indicated
L
In addition the measured temperature value
>
Ct 3
< is saved during the next step.
•
Integer digits flash, set by pressing
+
•
Decimal digits flash, set by pressing
+