Initial operation
21
7 Initial operation
7.1 Connect the device to the power supply
This section describes how the circulator is connected as a bridge mounted
circulator or heating circulator.
The circulator is mounted as a bridge mounted or heating circulator.
The power cable is ready for use.
1. Insert the power cable on the back of the circulator into the mains
connection [1].
2. Connect the circulator to the power supply using the power cable.
The circulator is connected.
7.2 Fill device
This section describes how the device should be filled with bath fluid during
initial operation.
Specifications for filling volume can be found in the technical data.
The device is mounted on a bath and switched off.
1. Fill the bath with water.
Maximum filling height 30 mm below the upper edge of the bath.
The bath fluid expands with increasing temperature and can overflow.
With decreasing temperature, the low liquid level protection can be
triggered and interrupt the temperature control process.
2. Place the sample in the bath.
3. If necessary, adjust the level by refilling or draining.