9.
Troubleshooting guide / Error messages
+
8
Whenever the microprocessor electronics registers a failure, a complete
shutdown of the compressor and circulating pump is performed. The alarm light
"
" illuminates and a continuous signal tone sounds.
The LED temperature display indicates the cause for the alarm in form of a
code.
Press enter
to quit the audible signal
.
x
The recirculating cooler is operated without bath fluid, or the liquid level is
insufficient.
Replenish the bath tank with the bath fluid.
x
Tube breakage has occured (insufficient filling level due to excessive bath
fluid pumped out). Replace the tubing and replenish the bath tank with the
bath fluid.
Cable of the working temperature sensor interrupted or short-circuited.
Error in A/D converter
The return temperature is above the switch-off value of the high temperature
protection (85°C). Check dimensioning of application.
Use a stronger recirculating cooler if necessary.
After eliminating the malfunction, press the mains power switch off and on
again to cancel the alarm state.
If the unit cannot be returned to operation, contact an authorized service
station.
Warning without a complete shutdown of the unit
Excess temperature warning starting at 82 °C
The return temperature soon reaches the swith-off value of the high
temperature protection (85 °C).
Cooling of the condenser is affected. (see page 31)
x
Clean air-cooled condenser.
x
Check the flow rate and cooling water temperature on water-cooled
condenser.
If the unit cannot be returned to operation, contact an authorized JULABO
service station.