FP51-SL
39
9.1.5.
Interruption after a power failure
It is possible to operate the programmer safely in case of a power
failure.
Important:
For this, >
Autostart<
needs to be set in the >
Configuration
< menu.
L
If the AUTOSTART function is activated, the programmer starts
again at a point approx. 20 seconds before the interruption took
place. However, an uncontrolled change of the bath temperature has
occurred.
L
Power failure when the instrument is on hold if start is effected via
the built-in timer:
If the starting time is not yet reached, the programmer is put on hold
again.
If the starting time is exceeded, the programmer starts immediately.
9.1.6.
Termination of a program
L
A program can be terminated by pressing
. The programmer
switches back to the Start menu.
Further procedures are defined in the MENU >START PROGRAM<,
menu item >Program End <.
Example: >Program End - StdBy<
The VFD COMFORT-DISPLAY displays “OFF”.
Interruption after a power failure
In case of a power failure, the reaction of the circulator is the same as
when switched off and on again with the mains switch.
The VFD COMFORT-DISPLAY displays “OFF”.