INSTRUCTION MANUAL CODE 80305 REV 0
Feb/2015 (It can be modi
fi
ed without notice) Page: 18
J
.P. SELECTA s.a.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
e-mail: [email protected] - website: http://www.jpselecta.es
7.3 TASK screen
There are two operation modes:
• Working at a unique temperature.
• Working according to a pre-programmed temperature pro
fi
le.
Press on to change the operation mode.
7.3.1 Operation at a temperature
• Select the set point temperature by pressing on the temperature indica-
tor.
• Select a temperature increase ramp if you want to increase it in a slowly
way.
The value 0ºC/min makes that the ramps increases as fast as possible.
• Select an operating time. This time starts to count down when the tem-
perature reaches the set point.
• The 0ºC value sets a continuous operation.
• Start by pressing on this icon:
7.3.2 Operation with a temperature pro
fi
le
• Temperature pro
fi
les must be previously programmed from the «PRO-
GRAMMATION» display.
• Select a pro
fi
le (from 0 to 9) by pressing on keys
• Start by pressing on the icon:
• In the temperature box (top-left corner) the current temperature and the
set point are indicated.
Fig 7.3.1. Run screen operating at one tem-
perature.
Fig 7.3. Run screen operating with tempera-
ture pro
fi
le.
95ºC
100ºC
Ramp
2ºC/min
01:00
01:00
100ºC
PROG 0
RAMP
Time 01:00
95ºC
Step: 2
2ºC/m
00:53
75ºC
01:00
2
08:00
100ºC
03:00
OFF
180ºC
INF
OFF
100ºC
03:00
OFF
100ºC
03:00
OFF
02:13
Red: Step number
Yelow: Waiting.
Verde:
Reading stored.
Blanco:
Reading not stored.
100ºC
PROG 2
RAMP
Time 01:00
95ºC
Step: 1
2ºC/m
00:53
Temperature
actual.
Nº step on
running.
Last ramp run.
Step temperature
set.
Step elapsed time.
Step set
time.
Fig 7.3.2 Working screen.
Temperature
set.
Internal oven tempe-