SAFETY THERMOSTAT:
To adjust the safety thermostat (10) temperature do as follows:
1. Turn the safety thermostat control knob (10) clockwise to the end
of the scale.
2. Select the desired safety temperature with the electronic regulator.
3. When the temperature is stabilized, turn the safety thermostat
control knob (10) slowly to the left until it clicks, the red indicator
lamp (7) lights up.
4. Stop the incubator with the main switch 1 and leave it to cool.
5. Start the apparatus and with the regulator select the working
temperature (minimum at 5ºC below the previous selected tem-
perature).
The safety thermostat is useful when the user works habitually at the
same temperature.
MAINTENANCE
Before removing the cover disconnect the equipment from the
mains.
The manipulation of the internal electronic circuits of the
equipment by unauthorized personnel can cause irreparable
damage. Take it to one of the J.P.SELECTA authorized technical
services.
CLEANING:
For the cleaning of the different parts use the following products:
Cleaning of stainless steel: Alcohol
Cleaning of plastic: Alcohol with cotton duster.
SPARE PARTS
To guarantee the safety of the equipment, all spare parts must
be purchased from J.P.SELECTA, s.a.u.
Descripción / Description
Código / Code
Regulador de temperatura /
Temperature controller
16123
Relé estático /
Static relay
36027
Motor /
Motor
24167
Termostato de seguridad /
Safety Thermostat
43031
Sonda Pt100 /
Pt100 probe
43059
Resistencia 3000W /
Heating element 3000W
39018