JOYTECH DET-3024b Owner'S Manual Download Page 6

22

19

21

20

Specifications

Care And Cleaning

Memory Mode

1.  The Memory Mode can be accessed either in forehead mode or 

     object mode:

     When the thermometer has been turned on and followed by 

     Figure 2/  or finished testing, press and hold the 

 Button 

     The letter M will appear in the 

 corner of the display.  

     (See Figure  )

8

Power

lower left

11

2.  The thermometer will automatically memorize the last 30

     temperature readings. 

Each time the 

 Button

     

is pressed, the screen displays past readings that correspond

     with a number 1-30. The number 1 reflects the most recent

     reading, while the number 30 reveals the oldest reading stored

     in memory.(See  Figure 12)   

Each memory also records the

     

measurement mode icon. 

Power

3.  

The user can press the 

Test Button 

to take new measurements. 

In the memory mode,      mark or      mark will not change.

     

 

Figure 12

Figure 11

Battery Replacement

1. 
     continue to measure; replace battery when "         " 

 in the 

 

When         flashes, it indicates that the power is low, but you can 

remains 

     displayed

upper

 right corner  of  LCD display as

 

shown

 

   

in

 

Figure 13.

2.  Slide battery cover down as shown in Figure  .

14

3.  Remove battery and install 2 new AAA alkaline batteries 

     as shown 

 . 

in Figure 

15

4.  Slide battery cover back on.

Figure 

13

Figure 

14

Figure 

15

Forehead mode: 34.0

~43.0

(93.2

~109.4

)

Object mode: 0

~100

(32

~212

)

Forehead mode: 

±

0.2

 (0.4

) during 35.5

~42.0

 (95.9

~107.6

at 15

~35

 (59.0

~95.0

) operating temperature range

±

0.3

 (0.5

) for other measuring and operating 

temperature range
Object mode: 

±

4% or 

±

2

(4

) whichever is greater

5

~40

(41

~104

), 15%~85%RH, non-condensing

Atmospheric Pressure : 70kPa~ 106kPa

Approx.85 grams(with batteries)

DC3V(2

×

AAA

 

battery)

Measuring range

Laboratory accuracy

Display resolution

0.1

 or 0.1

Measure time

Approximately 1 second

Operating temperature 
range:

Storage and transport 
temperature range

-25

~ 55

 (-13

~131

), 15%~95%RH, non-condensing

70kPa~ 106kPa

Atmospheric Pressure : 

Clinical accuracy

Dimension
Weight
Battery
Battery life

Approx. 3000 readings 

Measuring site    

        

Forehead(Forehead Mode)

 Oral (This thermometer converts the forehead temperature
 to display its “oral equivalent.”)

Reference body site

Operation mode                  Forehead mode(Adjust mode)

Ingress protecting rating
         
    

 IP22

Shock

withstands drop of 3 feet

Expected service life

Three years 

Contraindication

         
    

No contraindication

129.5*36*30.4mm

0-1 year:

Clinical bias: -0.17

℃( -

0.31

℉) ;

Clinical repeatability: 

0.06

℃(

0.11

℉) ;

Limits of agreement: 0.24

℃(

0.43

℉)

1

-

5

 

years:

Clinical bias: -0.17

℃( -

0.31

℉) ;

Clinical repeatability: 

0.07

℃(

0.13

℉) ;

Limits of agreement: 0.22

℃(

0.40

℉)

over

 

5

 

years

:

Clinical bias: -0.17

℃( -

0.31

℉) ;

Clinical repeatability: 

0.20

℃(

0.36

℉) ;

Limits of agreement: 0.26

℃(

0.47

℉)

1.  The probe window must be kept clean, dry, and undamaged at all

     times to ensure accurate readings. The accuracy of temperature 

     readings can be affected by damage to the probe window, or the 

     presence of dirt, fingerprints,dust and other soiling 

     compounds on the probe window. Degraded sensors can degrade

     performance or cause other problems.

 

2

For cleaning

       

1) Soak a clean soft cloth in drinking water,wring it out,and then 

          wipe the thermometer(including probe) no less than 3 times

       

2) Visual or use magnifying glass to observe the thermometer have

          no visible dirt and stains ,then use another clean soft cloth to 

          wipe the thermometer residue water

       

3) Put the thermometer in the original packaging.

  3.For disinfection

       

1) Soak a clean soft cloth in drinking water,wring it out,and then 

          wipe the thermometer(including probe) no less than 3 times

       

2) Visual or use magnifying glass to observe the thermometer have 

          no visible dirt and stains ,then use another clean soft cloth to 

          wipe the thermometer residue water

       

3) Using a clean soft cloth dipped in 70% medical alcohol,wipe the 

           probe for 3 Times,each time 1 minute.

      4) Using a clean cotton swab dipped in 70% medical alcohol,wipe 

           the sensor window 3 times

       

5) Wait at least 10 minutes to let the alcohol volatilization and put it 

           in the original packaging.  

4.  Do not put the thermometer into water directly.

5.  Store the thermometer in a dry location, free from dust and 

     contamination and away from direct sunlight.

6.  

Put the thermometer back to the original packaging after using. 

Summary of Contents for DET-3024b

Page 1: ...requirements for basic safety and essential performance of clinical thermometers for body temperature measurement IEC 60601 1 11 Medical electrical equipment Part 1 11 General requirements for basic...

Page 2: ...g to learn how to change between Chinese and English 1 Probe 2 Test Button 3 4 Setting Button 5 Battery Cover 6 Power Switch Display 1 2 3 4 5 6 forehead thermometer 2 Do not touch the temperature pro...

Page 3: ...ne hour until the correct hour appears After the hour is set press Button the Minute figure is flashing automatically Power Setting Set the time format The device can display the time in either an AM...

Page 4: ...st 10 minutes before taking a reading 3 Users should not drink eat or be physically active such as bathing showering shampooing and hair drying before while taking the measurement Remove hat and hair...

Page 5: ...ean the probe after each use to ensure an accurate reading and avoid cross contamination See the section of Care and Cleaning for details 30 seconds BackLight In Forehead mode 1 The display will be li...

Page 6: ...site Forehead Forehead Mode Oral This thermometer converts the forehead temperature to display its oral equivalent Reference body site Operation mode Forehead mode Adjust mode Ingress protecting ratin...

Page 7: ...oubleshooting Caution 95 15 131o F 13o F Manufacturer Manufacturing Date Direct Current Batch Code Storage and Transportation Temperature Limit 13 131 25 55 TYPE BF APPLIED PART Refer to instruction m...

Page 8: ...it not expressly approved by the party responsible for compliance could void the user authority to operate the equipment Note This equipment has been tested and found to comply with the limits for a C...

Page 9: ...ct and is subject to special precautionary measures with regard to EMC which must be published in the instructions for use Portable and mobile HF communications equipment can affect the device Use of...

Page 10: ...frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m Table 3 continued 37 Electromagnetic Compatibility Information 38 Electromagnetic Compatibility Information WARNINGS This de...

Reviews: