Specifications
Care And Cleaning
Memory Mode
1. The Memory Mode can be accessed either in forehead mode or
object mode:
When the thermometer has been turned on and followed by
Figure 2/ or finished testing, press and hold the
Button
The letter M will appear in the
corner of the display.
(See Figure )
8
Power
.
lower left
11
2. The thermometer will automatically memorize the last 30
temperature readings.
Each time the
Button
is pressed, the screen displays past readings that correspond
with a number 1-30. The number 1 reflects the most recent
reading, while the number 30 reveals the oldest reading stored
in memory.(See Figure 12)
Each memory also records the
measurement mode icon.
Power
3.
The user can press the
Test Button
to take new measurements.
In the memory mode, mark or mark will not change.
Figure 12
Figure 11
Battery Replacement
1.
continue to measure; replace battery when " "
in the
When flashes, it indicates that the power is low, but you can
remains
displayed
upper
right corner of LCD display as
shown
in
Figure 13.
2. Slide battery cover down as shown in Figure .
14
3. Remove battery and install 2 new AAA alkaline batteries
as shown
.
in Figure
15
4. Slide battery cover back on.
Figure
13
Figure
14
Figure
15
Forehead mode: 34.0
℃
~43.0
℃
(93.2
℉
~109.4
℉
)
Object mode: 0
℃
~100
℃
(32
℉
~212
℉
)
Forehead mode:
±
0.2
℃
(0.4
℉
) during 35.5
℃
~42.0
℃
(95.9
℉
~107.6
℉
)
at 15
℃
~35
℃
(59.0
℉
~95.0
℉
) operating temperature range
±
0.3
℃
(0.5
℉
) for other measuring and operating
temperature range
Object mode:
±
4% or
±
2
℃
(4
℉
) whichever is greater
5
℃
~40
℃
(41
℉
~104
℉
), 15%~85%RH, non-condensing
Atmospheric Pressure : 70kPa~ 106kPa
Approx.85 grams(with batteries)
DC3V(2
×
AAA
battery)
Measuring range
Laboratory accuracy
Display resolution
0.1
℃
or 0.1
℉
Measure time
Approximately 1 second
Operating temperature
range:
Storage and transport
temperature range
-25
℃
~ 55
℃
(-13
℉
~131
℉
), 15%~95%RH, non-condensing
70kPa~ 106kPa
Atmospheric Pressure :
Clinical accuracy
Dimension
Weight
Battery
Battery life
Approx. 3000 readings
Measuring site
Forehead(Forehead Mode)
Oral (This thermometer converts the forehead temperature
to display its “oral equivalent.”)
Reference body site
Operation mode Forehead mode(Adjust mode)
Ingress protecting rating
IP22
Shock
withstands drop of 3 feet
Expected service life
Three years
Contraindication
No contraindication
129.5*36*30.4mm
0-1 year:
Clinical bias: -0.17
℃( -
0.31
℉) ;
Clinical repeatability:
0.06
℃(
0.11
℉) ;
Limits of agreement: 0.24
℃(
0.43
℉)
1
-
5
years:
Clinical bias: -0.17
℃( -
0.31
℉) ;
Clinical repeatability:
0.07
℃(
0.13
℉) ;
Limits of agreement: 0.22
℃(
0.40
℉)
over
5
years
:
Clinical bias: -0.17
℃( -
0.31
℉) ;
Clinical repeatability:
0.20
℃(
0.36
℉) ;
Limits of agreement: 0.26
℃(
0.47
℉)
1. The probe window must be kept clean, dry, and undamaged at all
times to ensure accurate readings. The accuracy of temperature
readings can be affected by damage to the probe window, or the
presence of dirt, fingerprints,dust and other soiling
compounds on the probe window. Degraded sensors can degrade
performance or cause other problems.
2
.
For cleaning
:
1) Soak a clean soft cloth in drinking water,wring it out,and then
wipe the thermometer(including probe) no less than 3 times
;
2) Visual or use magnifying glass to observe the thermometer have
no visible dirt and stains ,then use another clean soft cloth to
wipe the thermometer residue water
;
3) Put the thermometer in the original packaging.
3.For disinfection
:
1) Soak a clean soft cloth in drinking water,wring it out,and then
wipe the thermometer(including probe) no less than 3 times
;
2) Visual or use magnifying glass to observe the thermometer have
no visible dirt and stains ,then use another clean soft cloth to
wipe the thermometer residue water
;
3) Using a clean soft cloth dipped in 70% medical alcohol,wipe the
probe for 3 Times,each time 1 minute.
4) Using a clean cotton swab dipped in 70% medical alcohol,wipe
the sensor window 3 times
;
5) Wait at least 10 minutes to let the alcohol volatilization and put it
in the original packaging.
4. Do not put the thermometer into water directly.
5. Store the thermometer in a dry location, free from dust and
contamination and away from direct sunlight.
6.
Put the thermometer back to the original packaging after using.