Troubleshooting
Problem
Solution
The ambient temperature
is not within the range
between 10
℃
and 40
℃
(50
℉
~
104
℉
).
Place the
in a room for at least 30
minutes at room
temperature between
10
℃
and 40
℃
(50
℉
~
104
℉
)
thermometer
The thermometer is not
functioning properly.
Unload the battery,
wait for 1 minute and
repower it. If the
message reappears,
contact the retailer
for service.
Error message
20
19
21
22
Troubleshooting
In Forehead mode:
Temperature taken is
higher than 43.0
℃
(109.4
℉
).
In Object mode:
Temperature taken is
higher than 100
℃
(212
℉
).
Problem
Solution
The thermometer could
not
work due to low
battery.
Replace two new
alkaline batteries
size AAA.
Error message
In Forehead mode:
Temperature taken is
lower than 34.0
℃
(93.2
℉
).
In Object mode:
Temperature taken is
lower than 0
℃
(32
℉
).
Read Temperature Taking
Hints Thoroughly, then
take a new temperature
measurement.
Read Temperature
Taking Hints thoroughly ,
then make sure the lens
filter are clean, then take
a new temperature
measurement.
Specifications
Forehead mode: 34.0
℃
~43.0
℃
(93.2
℉
~109.4
℉
)
Object mode: 0
℃
~100
℃
(32
℉
~212
℉
)
Forehead mode:
±
0.2
℃
(0.4
℉
) during 35.5
℃
~42.0
℃
(95.9
℉
~107.6
℉
)
at 15
℃
~35
℃
(59.0
℉
~95.0
℉
) operating temperature range
±
0.3
℃
(0.5
℉
) for other measuring and operating
temperature range
Object mode:
±
4% or
±
2
℃
(4
℉
) whichever is greater
41*26.5*153mm
Approx.84grams(with batteries)
DC3V(2
×
AAA
battery)
Measuring range
Laboratory accuracy
Display resolution
0.1
℃
or 0.1
℉
Measure time
1 second
Operating temperature
range:
Storage and transport
temperature range
-25
℃
~ 55
℃
(-13
℉
~131
℉
), 15%~95%RH, non-condensing
700hPa ~ 1060hPa
Atmospheric Pressure :
Clinical accuracy
Dimension
Weight
Battery
Battery life
Approx. 1 year/6000 readings
Measuring site
Forehead(Forehead Mode)
Oral (This thermometer converts the forehead temperature
to display its “oral equivalent.”)
Reference body site
Operation mode Forehead mode(Adjust mode)
Ingress protecting rating
IP22
Shock
withstands drop of 3 feet
Expected service life
Three years
Clinical bias: -0.09
℃(-
0.16
℉)
Clinical repeatability: 0.13
℃(
0.23
℉)
Limits of agreement: 0.87
℃(
1.57
℉)
10
℃
~40
℃
(50
℉
~104
℉
), 15%~85%RH, non-condensing
700hPa ~ 1060hPa
Atmospheric Pressure :
Calibration
The thermometer is initially calibrated at the time of manufacture.
If the thermometer is used according to the use instruction, periodic
readjustment is not required. However, We recommends checking
calibration every two years or whenever clinical accuracy of the
thermometer is in question. Please send the complete device to the
dealers or manufacturer.
The above recommendations do not supersede the legal requirements.
The user must always comply with legal requirements for the control
of the measurement, functionality, and accuracy of the device which
are required by the scope of relevant laws, directives or ordinances
where the device is used.
A clinical summary and procedures for checking calibration are
available upon request.(Turn on the thermometer and press the power
button long time until entering into calibrate mode, software
version
will be displayed.)
Direct Current
Disposal of this product and used batteries should be carried out in accordance
with the national regulations for the disposal of electronic products
Type B F applied part
Consult Accompanying Documents
LOT
Batch Code
Storage and Transportation Temperature Limit:
-13
℉ ~
131
℉
(-25
℃
~55
℃)
-25
℃
55
℃
1. Probe
2. START Button
3. Set Button
4. Battery Cover