Electromagnetic Compatibility Information

NOTE 1At 80 MHz and 800 MHz, the higher frequency range applies.NOTE 2 
These guidelines may not apply in all situations. Electromagnetic propagation is 
affected by absorption and reflection from structures, objects and people.

a The ISM(industrial, scientific and medical) bands between 0,15 MHz and 80 MHz 
are 6,765 MHzto6,795 MHz;13,553 MHz to 13,567 MHZ;26,957 MHz to 27,283 
MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz 
and 80 MHz are 1,8 MHz to 2,0 MHz,3,5 MHz to 4,0 MHz,5,3 MHz to 5,4 MHz,7 
MHz to 7,3 MHz,10,1 MHz to 10,15 MHz,14 MHz to 14,2 MHz,18,07 MHz to 18,17 
MHz,21,0MHz to 21,4MHz,24,89 MHz to 24,99 MHz,28,0 MHz to 29,7 MHz and 
50,0 MHz to 54,0 MHZ.

b The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz 
and in the frequency range 80 MHz to 2,7 GHz are intended to decrease the 
likelihood that mobile/portable communications equipment could cause 
interference if it is inadvertently brought into patient areas. For this reason, an 
additional factor of 10/3 has been incorporated into the formulae used in 
calculating the recommended separation distance for transmitters in these 
frequency ranges.

c Field strengths from fixed transmitters, such as base stations for radio(cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio 
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To 
assess the electromagnetic environment due to fixed RF transmitters, an 
electromagnetic site survey should be considered. If the measured field strength in 
the location in which the device is used exceeds the applicable RF compliance level 
above, the device should be observed to verify normal operation. If abnormal 
performance is observed, additional measures may be necessary, such as 
re-orienting or relocating the device.

d  Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 
3 V/m.

Table 3 continued

Electromagnetic Compatibility Information

Electromagnetic Compatibility Information

WARNINGS!
· This device should not be used in the vicinity or on the top of other
  electronic equipment such as cell phone, transceiver or radio 
  control products. If you have to do so, the device should be 
  observed to verify normal operation.
· The use of accessories and power cord other than those specified, 
  with the exception of cables sold by the manufacturer of the 
 equipment or system as replacement parts for internal components, 
 may result in increased emissions or decreased immunity of the 
 equipment or system.

37

Electromagnetic Compatibility Information

38

14

Summary of Contents for DET-1013b

Page 1: ...e ear canal of human It converts the measured heat into a temperature reading displayed on the LCD When properly used it will quickly assess your temperature in an accurate manner The infrared thermom...

Page 2: ...e clock setting section to learn how to setup the time in the first use The thermometer has been designed for practical use It s not meant to replace a visit to the doctor Please also remember to comp...

Page 3: ...the unit Press Setting Button to select the unit you want After the unit is set press Mode Button the time format figure will appear Set the time format The device can display the time in either an AM...

Page 4: ...ver on the probe firmly making sure the probe and the inner portion of the probe cover fit snugly 2 Press the The display is activated to show all segments Remarks a Probe cover sign stand for taking...

Page 5: ...e Taking Hints 18 Temperature Taking Hints Figure 5 Figure 6 Bluetooth requirements The thermometer requires a bluetooth device with Bluetooth 4 0 or later Android 6 0 or later IOS 8 0 or later And wo...

Page 6: ...mometer and probe cover in a dry location free from dust and contamination and away from direct sunlight 8 Periodic cleaning and disinfection of the device following use to prevent patient cross infec...

Page 7: ...nctioning properly Read Illustration For Use thoroughly and take a new temperature measurement Replace two new alkaline batteries size AAA Allow the thermometer to rest in a room for at least 30 minut...

Page 8: ...amage during shipping Include the original sales slip indicating the date of purchase a note describing the problem and your return address Send the thermometer prepaid and insured The lay operator or...

Page 9: ...The device is an electrical medical product and is subject to special precautionary measures with regard to EMC which must be published in the instructions for use Portable and mobile HF communicatio...

Page 10: ...c Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predic...

Reviews: