A.
)
Menu Button- allows you to change internal settings
use arrows on left side of keypad to
maneuver through.
I.
_
FILTER SETTINGS
0.05 - 150 Hz No filter is selected
H50 D – Only the A/C filter and baseline drift filter are turned on
H50 D 75 Hz - All filters are turned on and the frequency of muscle electricity filter is 0.5 -75Hz
H50 D 35 Hz – All filters are turned on and the frequency of muscle electricity filter is 0.5 -35Hz
II. BASE LINE WIDTH =
LINE SIZE OF BASE LINE on ECG printout.
1X- Thinnest
2X - Default
3X
4X- Thickest
III. EXTERNAL RECORD
-
Turn on / off record of external input
Signal, this should be set to
off
.
IV. LCD Contrast-
Adjusts contrast on LCD screen from 1-16, default =7 .
V. LAUNGAGE SET
-
Chinese, English, Russian
Default is English.
VI. AC FREQUENCY
–
Setting the A/C frequency
Default should be set to 50 Hz.
VI. PRINTER TEST
-
Tests printer function.
press OK to test – prints triangular form
press Ok again to stop.
VII. RESTORE SET
-
Resets
f
actory default settings press OK to
ACTIVATE
VIII. RHYTHM LEAD
-
Selects the rhythm lead
IX. RECORD LENGTH
-
Length of time each lead will print in the auto
mode, ranges from 3-12 seconds.
7