Handheld Optical Light Source-JW3116
12
5.5 Wavelength Automatic Identification
⚫
Connect JW3116 light source with its same serials power meter JW3216.
⚫
Enable light source be Under “Wave ID” operation mode: press
or
of light source to
laser on JW2116, Hold down
for few seconds
,
light source will be enter into Wave ID mode,
also “--AU” will be shown on the upper right of LCD for a indication.
⚫
Enable power meter be under “Wave ID”operation mode: Hold down
for few seconds
,
power
meter will be enter into Wave ID mode
, also “--AU”will be shown on the upper right of LCD for a
indication.
⚫
Once the ID information is changed from light source(press
to change wavelength), after 3 to
5 seconds later, the detected information on optical power meter also will be changed automatically
according to light source. Please refer to below figure(5-14) for easy understanding.
⚫
Exit Wave ID mode: Hold down
again to exit Wave ID mode from power meter and hold
down
again to exit Wave ID mode from light source.
JW3116 Light source JW3216 Power Meter
5-14
5.6 Frequency detection
⚫
Connect JW3116 light source with its same serials power meter JW3216.
⚫
Output frequency from JW3116 optical light source: Press
or
to emerge light
from the unit, press
for very short second
,
JW3116 light source will output frequencies
of 270Hz
、
330Hz
、
1KHz
、
2KHz accordingly
,
which will be shown on the upper right of the LCD
in JW3116 light source. In the mean time, the JW3216 power meter will detect the
corresponding frequency automatically from JW3116 light source. Please refer to below figure
(5-15) for better understanding:
1310nm(xxdBm)
1550nm(xxdBm)
……
1310nm(xxdBm)
1550nm(xxdBm)
……