10 Hints for Radiation Contamination Monitoring
Hold detector probe 2mm from the surface being measured. This provides conditions which most
closely reflect those used when calibrating the probe.
Avoid contaminating the face or head of the probe during operation.
Wherever possible, check that the monitor is working before use by placing it adjacent to a known
test source.
Alph a part icles are h e avil y a tte nu ate d b y wa ter. Mo n itor in g a wet surf ace containing
alpha emitting radionuclides will dramatically reduce the observed count rate relative to dr y
conditions. In the worst case scenario, all alpha particles will be absorbed before being
detected by the probe. A similar, though less pronounced problem may be encountered when
measuring nuclides which emit weak beta particles.
Use the optional extension arm for the detector probe when monitoring above head height
or the internal surface of pipes, vessels or ventilation ducts.