AD-1272 Advanced Thermal Dispersion Probe Airflow Measuring System Technical Bulletin
26
3. Press
ENTER
. The display indicates the current K-Factor setting (Off or On).
4. Press
UP or DOWN
to change the setting and
ENTER
to confirm the setting. The new setting is stored in
memory and the display returns to the K-Factor Configuration submenu.
5. Press
UP or DOWN
to scroll to the
Calculate K-Factor
submenu selection and press
ENTER
.
6. Press
UP or DOWN
to scroll between Yes or No. Press
ENTER
to make a selection. If No is selected, see
Automatic Calculation of K-Factor Not Enabled
Automatic Calculation of K-Factor
Automatic Calculation of K-Factor Not Enabled
1. When K-Factor is not enabled, the K-Factor Configuration submenu formats itself for manual K-Factor entry
mode. Scroll to the
K-Factor Gain
submenu selection and press
ENTER
.
2. Press
ENTER
to adjust the manual K-Factor gain. Press
UP or DOWN
to modify the currently selected value.
3. Press
ENTER
to confirm the value and return to the K-Factor Configuration submenu.
Figure 40: K-Factor Settings Screen
Figure 41: Calculate K-Factor Selection Screen
Figure 42: K-Factor Gain Selection Screen
Figure 43: K-Factor Gain Screen