Program Settings
Z370584 Rev. D
6-8
Program Settings
Phaco
1.
Tap
to open the Settings screen.
2. Select a Phaco submode on the left (
).
Figure 6-6 — Phaco Settings Screen
3. If desired, tap the submode name at the top of the screen (outlined in red in
customize the submode name.
4. Select the pump type (
Peristaltic
or
Venturi
).
5. Use the up and down arrows to set the
Aspiration
rate (10 to 60).
6. Use the up and down arrows to set the
Vacuum
. If you are using Venturi, the range is 0 to
600. If you are using Peristaltic, the range is 0 to 650.
7. Use the up and down arrows to set the
Power
(0% to 100%).
8. Select the delivery mode (
Linear
or
Panel
).
9. Set the
Bottle Height
(Advanced Fluidics) or
Total Height
(Advanced Infusion).
10. Tap
Save
to save the program.
For additional information see
“FUSION Fluidics Phaco” on page 6-27
and
Summary of Contents for Veritas
Page 2: ...Z370584 Rev D This page was intentionally left blank...
Page 4: ...Z370584 Rev D This page was intentionally left blank...
Page 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Page 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Page 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Page 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Page 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Page 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Page 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Page 270: ...Z370584 Rev D I 8 This page was intentionally left blank...