Surgeon Settings
Z370584 Rev. D
Foot Pedal Setup
5-23
3. Press the foot pedal down fully to position 3. As the foot pedal travels through positions 1
and 2 to position 3, the progress bar shows its current state. When the pedal reaches
position 3, the progress bar should be solid green.
4. Tap each of the four switches. When a switch is pressed, the On button for that switch
turns green and its Off button turns white. When you release the switch, the On button
for that switch turns white and its Off button turns green.
5. Tap the
Finished
button when you complete the testing of your foot switch.
6. Tap
Exit
to close the Foot Pedal Set Up screen.
Advanced Control Pedal Setup
This section presents the following topics for setting up an Advanced Control Pedal (ACP):
•
“Vitrectomy Foot Pedal Modes” on page 5-24
•
“Advanced Control Pedal Connections” on page 5-24
•
“Advanced Control Pedal Switch Assignment” on page 5-24
•
“Advanced Control Pedal Threshold Configuration” on page 5-28
•
“Advanced Control Pedal Calibration and Testing” on page 5-30
Figure 5-20 — Advanced Control Pedal
1. Left yaw
4. Heel rest
7. Right yaw
10. Wireless Indicator
2. Wake-up button
5. Right heel switch
8. Handle
11. Battery Indicator
3. Left heel switch
6. Lights
9. Treadle
Summary of Contents for Veritas
Page 2: ...Z370584 Rev D This page was intentionally left blank...
Page 4: ...Z370584 Rev D This page was intentionally left blank...
Page 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Page 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Page 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Page 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Page 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Page 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Page 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Page 270: ...Z370584 Rev D I 8 This page was intentionally left blank...